Establishment of regional diagnostic reference levels for digital mammography in Western Province of Sri Lanka.

J Radiol Prot

Department of Radiological Sciences, Graduate school of Human Health Sciences, Tokyo Metropolitan University, Tokyo, Japan.

Published: February 2021

The radiation dose to the breasts should be kept to a minimum as breast tissues are highly sensitive to radiation. In mammography, the mean glandular dose (MGD) is used to specify the patient dose. In this study, data on the MGD during diagnostic mammographic examinations was collected using the database from six digital mammography facilities available in the Western Province in Sri Lanka. Examinations involving breast pathology, breast implants, or compressed breast thicknesses (CBT) outside the range of 20-110 mm were excluded in this study. The mean MGD per breast was 3.50 mGy, with a mean CBT of 57 mm. The mean MGD per facility varies from 1.58 to 2.27 mGy, with overall 75th and 95th percentiles of 2.15 and 2.82 mGy, respectively. The 75th and 95th percentile MGD per image, for the average CBT of 57 ± 12 mm, were 2.00 and 2.65 mGy respectively. The 75th percentile value of the MGD is suggested for the Western Province and it depends on the specific CBT.

Download full-text PDF

Source
http://dx.doi.org/10.1088/1361-6498/abd075DOI Listing

Publication Analysis

Top Keywords

western province
12
digital mammography
8
province sri
8
sri lanka
8
75th 95th
8
percentile mgd
8
mgd
6
breast
5
establishment regional
4
regional diagnostic
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!