Background: The Mindray BC-6200 is a new automatic hematology analyzer that quantifies the parameters of blood morphology and leukocyte differential in five populations (5-Diff). The aim of the study was to evaluate the BC-6200 and compare it with the Siemens ADVIA 2120i analyzer.

Materials And Methods: The comparison between BC-6200 and ADVIA 2120i analyzers was performed using 390 whole blood samples collected on K EDTA. For the BC-6200, the carryover effect, precision, and linearity were evaluated. 138 samples were used to assess the sensitivity and flag ability, suggesting the presence of abnormal cells such as blasts, immature granulocytes, or atypical lymphocytes. Flagging results were compared with microscopic evaluation of blood smears.

Results: The BC-6200 analyzer showed a high correlation (r ≥ .97) with ADVIA 2120i for most of the compared parameters except RDW (r = .8350), MPV (r = .7634), Mon# (r = .8366), Baso# (r = .9205), and NRBC (r = .3768). The BC-6200 had better correlation with microscopic evaluation for NRBC (r = .8902) compared with ADVIA 2120i (r = .5677). The BC-6200 has shown high efficiency for flagging blasts (80.4%), immature granulocytes (80.5%), and atypical lymphocytes (69.0%).

Conclusion: The new Mindray BC-6200 hematology analyzer provides high measurements precision and good correlation with ADVIA 2120i for most of the morphology and 5-diff parameters.

Download full-text PDF

Source
http://dx.doi.org/10.1111/ijlh.13418DOI Listing

Publication Analysis

Top Keywords

advia 2120i
24
mindray bc-6200
12
hematology analyzer
12
bc-6200
9
bc-6200 hematology
8
immature granulocytes
8
atypical lymphocytes
8
microscopic evaluation
8
analyzer high
8
advia
6

Similar Publications

Aim: Our goal was to examine the effect of high-intensity physical activity on changes in the lipid profile, complete blood count (CBC), iron metabolism, and kidney and liver function tests of professional water polo players.

Methods: This study included twenty professional male water polo players. Blood sampling was carried out at the beginning of the season and during periods of high-intensity training.

View Article and Find Full Text PDF

A comparative study of blood cell count in four automated hematology analyzers: An evaluation of the impact of preanalytical factors.

PLoS One

May 2024

Translational Science & Experimental Medicine, Research and Early Development, Respiratory & Immunology, BioPharmaceuticals R&D, AstraZeneca, Gaithersburg, MD, United States of America.

Article Synopsis
  • Differential white blood cell counts are important for diagnosing and treating conditions like asthma, but different testing methods can yield variable results under different conditions.
  • A study compared blood samples analyzed with four popular hematology analyzers to assess the stability of eosinophils, a key cell type in asthma management, under various storage temperatures and time points.
  • Findings indicated that while fresh samples provided similar results across platforms, significant discrepancies arose after 24 hours, particularly with eosinophils, emphasizing the need for optimized handling conditions based on cell type and analyzer used.
View Article and Find Full Text PDF

Introduction If non-odontogenic abscesses and phlegmons (all purulent inflammations where the etiology is not a diseased tooth) of the neck are not promptly treated, they can lead to serious complications and even end in the death of the affected patient. Classical markers of inflammation such as plasma concentration of leukocytes (WBC), neutrophils (Neu), C-reactive protein (CRP), and erythrocyte sedimentation rate (ESR) are elevated in inflammatory neck diseases, but none of them has been proven as a definite marker in the prediction of this type of pathology. This necessitates the search and analysis of new indicators that could be used in the diagnosis, follow-up, and prognosis of patients with purulent neck infections.

View Article and Find Full Text PDF

Comparison of visual assessments of anisocytosis in canine blood smears and analyzer-calculated red blood cell distribution width.

Front Vet Sci

September 2023

Department of Veterinary Clinical Sciences, Faculty of Health and Medical Sciences, University of Copenhagen, Copenhagen, Denmark.

Red blood cell distribution width (RDW) and visual assessments of anisocytosis assess variability in erythrocyte size. Veterinary studies on the correlation between the two methods and on observer agreement are scarce. The objectives were to assess the correlation of the grading of anisocytosis by means of conventional microscopy of canine blood smears to RDW, and to assess intra- and inter-observer variation in assessing the degree of anisocytosis.

View Article and Find Full Text PDF

Background: Clozapine is a first-line therapy and the only FDA-approved drug for patients with treatment-resistant schizophrenia (TRS). However, frequent measurement of absolute neutrophil count (ANC) is required to monitor for potential adverse severe neutropenia from clozapine therapy. We evaluated 3 point-of-care (POC) instruments that perform the complete blood count (CBC) with differential to assess their analytical performance and potential to meet the clinical need for clozapine therapy management.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!