Objectives: To estimate the efficacy and safety of SuperPulsed thulium-fiber laser ureteral lithotripsy and to identify optimal laser settings.

Methods: Patients with solitary stones were prospectively included. Lithotripsy was performed with a SuperPulsed thulium-fiber laser (NTO IRE-Polus, Fryazino, Russia) using a rigid ureteroscope 7.5 Ch (Richard Wolf, Knittlingen, Germany). We analyzed the efficacy of lithotripsy by measuring total energy required for stone disintegration, "laser-on" time, ablation speed, ablation efficacy, and energy consumption. Stone retropulsion and visibility were assessed using a three-point Likert scale. Complications were assessed using the Clavien-Dindo classification system.

Results: A total of 149 patients were included. The mean stone density was 985 ± 360 Hounsfield units, the median (interquartile range) stone volume was 179 (94-357) mm . The median (interquartile range) total energy was 1 (0.4-2) kJ, and laser-on time 1.2 (0.5-2.7) min. The median (interquartile range) stone ablation speed was 140 (80-279) mm /min, energy for ablation of 1 mm was 5.6 (3-9.9) J/mm and energy consumption was 0.9 (0.6-1) J/min. A correlation was found between retropulsion and the energy used (r = 0.5, P < 0.001). Multivariable analysis showed energy to be a predictor of increased retropulsion (odds ratio 65.7, 95% confidence interval 1.6-2774.1; P = 0.028). No predictors for worse visibility were identified.

Conclusion: The SuperPulsed thulium-fiber laser provides effective and safe lithotripsy during ureteroscopy regardless of stone density. Fiber diameter and laser frequency do not influence visibility or safety. Optimal laser settings are 0.5 J × 30 Hz for fragmentation and 0.15 J × 100 Hz for dusting.

Download full-text PDF

Source
http://dx.doi.org/10.1111/iju.14443DOI Listing

Publication Analysis

Top Keywords

superpulsed thulium-fiber
12
thulium-fiber laser
12
median interquartile
12
interquartile range
12
laser ureteral
8
total energy
8
ablation speed
8
energy consumption
8
range stone
8
energy
6

Similar Publications

Article Synopsis
  • - This clinical study compares the effectiveness and safety of two lasers, the holmium: yttrium-aluminum-garnet (HO:YAG) and the superpulse thulium fiber laser (SP TFL), in treating kidney stones via ureterorenoscopic lithotripsy.
  • - Patients will be randomly assigned to receive treatment with either laser, with outcomes measured for stone elimination and various surgical performance indicators 30 days post-operation.
  • - The study is registered under ChiCTR2300076893 and aims to provide insights on whether the newer SP TFL can serve as a better alternative to the established HO:YAG laser.
View Article and Find Full Text PDF

Ablation efficiency and laser safety of a novel superpulsed thulium fiber laser: a in vitro study.

World J Urol

October 2024

Department of Urology, Tianjin Institute of Urology, The Second Hospital of Tianjin Medical University, 23 Pingjiang Road, Hexi, Tianjin, 300000, People's Republic of China.

Purpose: To assess the ablation efficiency of the Superpulsed Thulium Fiber Laser (SP-TFL) and investigate the thermal effects of SP-TFL.

Methods: A SP-TFLwas employed to evaluate ablation efficiency. Fresh ex-vivo pig kidneys and ureters were utilized to evaluate the renal pelvis and ureter temperature changes, different irrigation rates(0, 15, 38mL/min) and a long pulse width were used.

View Article and Find Full Text PDF

The super-pulsed thulium fiber laser (SP TFL) is a new alternative to high-power holmium laser for intracorporeal lithotripsy. The SP TFL has shown advantages in dusting regimes, but benefits in fragmentation regimes are less understood. The second-generation SP TFL introduces an advanced fragmentation pulse (AFP) sequence to maximize SP TFL's efficiency in fragmentation.

View Article and Find Full Text PDF

[Efficacy and safety of superpulse thulium laser lithotripsy in the intracavitary treatment of urinary calculi].

Zhonghua Yi Xue Za Zhi

August 2023

Department of Urology, Sun Yat-sen Memorial Hospital, Sun Yat-sen University, Guangzhou 510120, China.

To evaluate the efficacy and safety of superpulse thulium laser lithotripsy in the intracavitary treatment of urinary calculi. From May 2021 to July 2022, patients diagnosed with urinary calculi were screened in four medical centers. Those who met the criteria were treated with superpulse thulium fiber laser under endoscope.

View Article and Find Full Text PDF

Purpose: Holmium:YAG laser lithotripsy requires high amperage power and has an upper limit of frequency and a minimal fiber size. The technology utilizing thulium-doped fiber offers low pulse energy settings and high pulse frequencies up to 2,400 Hz. We compared the novel SuperPulsed thulium fiber laser (SOLTIVE™; Olympus) to a commercially available 120 W Ho:YAG laser.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!