Agreement between 2 swept-source OCT biometers and a Scheimpflug partial coherence interferometer.

J Cataract Refract Surg

From Oftalvist, Alicante, Spain (Tañá-Rivero, Aguilar-Córcoles, Tello-Elordi, Pastor-Pascual); Optics and Optometry and Vision Sciences Department, University of Valencia, Valencia, Spain (Montés-Micó).

Published: April 2021

Purpose: To evaluate the agreement between different parameters obtained with 2 swept-source optical coherence tomography (SS-OCT)-based biometers and 1 Scheimpflug camera with partial coherence interferometry (PCI).

Setting: Single center, Oftalvist, Alicante, Spain.

Design: Prospective case series.

Methods: Biometry was performed in 49 eyes using 3 optical biometers: ANTERION SS-OCT, IOLMaster 700 SS-OCT, and Pentacam AXL PCI. Keratometry (K), J0 and J45 vectors, anterior chamber depth (ACD), central corneal thickness (CCT), white-to-white (WTW), lens thickness (LT), and axial length (AL) were measured with each device. Bland-Altman analysis was applied.

Results: This study comprises 49 eyes of 49 patients. There were no statistically significant differences for K1, K2, J0 and J45 between the 3 devices (P > .9). In contrast, there was a statistically significant difference in the ACD, CCT, WTW, LT, and AL between the biometers (P < .001). Specifically, there was a statistically significant difference between ACD, CCT, and WTW values for all-pairwise comparisons. IOLMaster showed the shortest ACD value and ANTERION showed the largest ACD. IOLMaster showed the highest CCT and Pentacam showed the lowest CCT. IOLMaster showed the largest WTW and Pentacam showed the shortest WTW. The LT measured with IOLMaster was thicker than that measured with ANTERION. There was a statistically significant difference in the AL between IOLMaster and Pentacam, with a shorter AL measured with IOLMaster (P < .001), but no differences were found between ANTERION and IOLMaster (P = .599) and between ANTERION and Pentacam (P = .054).

Conclusions: Mean differences and the limits of agreement obtained in all-pairwise comparisons of the different parameters should be judged clinically to consider the interchangeability of these devices.

Download full-text PDF

Source
http://dx.doi.org/10.1097/j.jcrs.0000000000000483DOI Listing

Publication Analysis

Top Keywords

statistically difference
12
biometers scheimpflug
8
partial coherence
8
iolmaster
8
difference acd
8
acd cct
8
cct wtw
8
all-pairwise comparisons
8
measured iolmaster
8
anterion
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!