Purpose: UTE sequences typically acquire data during the ramping up of the gradient fields, which makes UTE imaging prone to eddy current and system delay effects. The purpose of this work was to use a simple gradient impulse response function (GIRF) measurement to estimate the real readout gradient waveform and to demonstrate that precise knowledge of the gradient waveform is important in the context of high-resolution UTE musculoskeletal imaging.

Methods: The GIRF was measured using the standard hardware of a 3 Tesla scanner and applied on 3D radial UTE data (TE: 0.14 ms). Experiments were performed on a phantom, in vivo on a healthy knee, and in vivo on patients with spine fractures. UTE images were reconstructed twice, first using the GIRF-corrected gradient waveforms and second using nominal-corrected waveforms, correcting for the low-pass filter characteristic of the gradient chain.

Results: Images reconstructed with the nominal-corrected gradient waveforms exhibited blurring and showed edge artifacts. The blurring and the edge artifacts were reduced when the GIRF-corrected gradient waveforms were used, as shown in single-UTE phantom scans and in vivo dual-UTE gradient-echo scans in the knee. Further, the importance of the GIRF-based correction was indicated in UTE images of the lumbar spine, where thin bone structures disappeared when the nominal correction was employed.

Conclusion: The presented GIRF-based trajectory correction method using standard scanner hardware can improve the quality of high-resolution UTE musculoskeletal imaging.

Download full-text PDF

Source
http://dx.doi.org/10.1002/mrm.28566DOI Listing

Publication Analysis

Top Keywords

high-resolution ute
12
gradient waveforms
12
gradient
9
trajectory correction
8
gradient impulse
8
impulse response
8
response function
8
ute
8
ute imaging
8
gradient waveform
8

Similar Publications

Pulmonary Structural MRI using Free-Breathing, Self-Gated Ultra-short Echo Time Imaging.

J Vis Exp

September 2024

Division of Pulmonary, Critical Care, and Sleep Medicine, Hoglund Biomedical Imaging Center, University of Kansas Medical Center;

High quality MRI of the lungs is challenged by low tissue density, fast MRI signal relaxation, and respiratory and cardiac motion. For these reasons, structural imaging of the lungs is performed almost exclusively using Computed Tomography (CT). However, CT imaging delivers ionizing radiation, and thus is less well suited for certain vulnerable populations (e.

View Article and Find Full Text PDF

Purpose: Ultrashort echo time (UTE) MRI can be a radiation-free alternative to CT for craniofacial imaging of pediatric patients. However, unlike CT, bone-specific MR imaging is limited by long scan times, relatively low spatial resolution, and a time-consuming bone segmentation workflow.

Methods: A rapid, high-resolution UTE technique for brain and skull imaging in conjunction with an automatic segmentation pipeline was developed.

View Article and Find Full Text PDF

Computed tomography (CT) is the preferred imaging modality for bone evaluation of the knee, while MRI of the bone is actively being developed. We present three techniques using short-interval delta ultrashort echo time (δUTE), field echo (FE), and FE with high resolution-deep learning reconstruction (HR-DLR) for direct bone MRI. Knees of healthy volunteers ( = 5, 3 females, 38 ± 17.

View Article and Find Full Text PDF

In-situ imaging of heat-induced phase transition in a two-dimensional conjugated metal-organic framework.

Micron

September 2024

Central Facility for Materials Science Electron Microscopy, Universität Ulm, Ulm 89081, Germany; Institute for Quantum Optics, Universität Ulm, Ulm 89081, Germany. Electronic address:

Atomically-resolved in-situ high-resolution transmission electron microscopy (HRTEM) imaging of the structural dynamics in organic materials remains a major challenge. This difficulty persists even with aberration-corrected instruments, as HRTEM images necessitate a high electron dose that is generally intolerable for organic materials. In this study, we report the in-situ HRTEM imaging of heat-induced structural dynamics in a benzenehexathiol-based two-dimensional conjugated metal-organic framework (2D c-MOF, i.

View Article and Find Full Text PDF

More accurate trabecular bone imaging using UTE MRI at the resonance frequency of fat.

Bone

July 2024

Department of Radiology, University of California, San Diego, La Jolla, CA, USA; Radiology Service, Veterans Affairs San Diego Healthcare System, San Diego, La Jolla, CA, USA. Electronic address:

High-resolution magnetic resonance imaging (HR-MRI) has been increasingly used to assess the trabecular bone structure. High susceptibility at the marrow/bone interface may significantly reduce the marrow's apparent transverse relaxation time (T2*), overestimating trabecular bone thickness. Ultrashort echo time MRI (UTE-MRI) can minimize the signal loss caused by susceptibility-induced T2* shortening.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!