Introduction: Low-kV IORT (Low kilovoltage intraoperative radiotherapy) using INTRABEAM machine and dedicated spherical applicators is a candidate modality for breast cancer treatment. The current study aims to quantify the RBE (relative biologic effectiveness) variations of emitted X-rays from the surface of different spherical applicators and bare probe through a hybrid Monte Carlo (MC) simulation approach.
Materials And Methods: A validated MC model of INTRABEAM machine and different applicator diameters, based on GEANT4 Toolkit, was employed for RBE evaluation. To doing so, scored X-ray energy spectra at the surface of each applicator diameter/bare probe were used to calculate the corresponding secondary electron energy spectra at various distances inside the water and breast tissue. Then, MCDS (Monte Carlo damage simulation) code was used to calculate the RBE values according to the calculated electron spectra.
Results: Presence of spherical applicators can increase the RBE of emitted X-rays from the bare probe by about 22.3%. In return, changing the applicator diameter has a minimal impact (about 3.2%) on RBE variation of emitted X-rays from each applicator surface. By increasing the distance from applicator surface, the RBE increments too, so that its value enhances by about 10% with moving from 2 to 10 mm distance. Calculated RBE values within the breast tissue were higher than those of water by about 4% maximum value.
Conclusion: Ball section of spherical IORT applicators can affect the RBE value of the emitted X-rays from INTRABEAM machine. Increased RBE of breast tissue can reduce the prescribed dose for breast irradiation if INTRABEAM machine has been calibrated inside the water.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.ejmp.2020.11.018 | DOI Listing |
Comput Methods Programs Biomed
September 2021
Physics Department, Hakim Sabzevari University, Daneshgah Blvd, P.O. 9617976487, Sabzevar, Iran. Electronic address:
Introduction: Intraoperative radiotherapy (IORT) by low energy X-rays is a single fraction treatment modality for tumor bed irradiation after breast-conserving surgery. It has been shown that the variations of breast tissue composition can affect the absorbed dose in this method. Apart from physical quantities such as absorbed dose value, radiobiological quantities including relative biological effectiveness (RBE) may also change with the variations of breast tissue composition.
View Article and Find Full Text PDFJ Biomed Phys Eng
February 2021
MD, Department of General Surgery, Mashhad University of Medical Sciences, Mashhad, Iran.
Background: Breast cancer is the most common cancer among women. Considering the fact that a high dose is delivered in a single fraction of IORT, the evaluation of the dose at sensitive organs like thyroid is necessary.
Objective: The current study has aimed to evaluate the received dose to thyroid lobes in the breast IORT technique.
Phys Med
December 2020
Physics Department, Hakim Sabzevari University, Daneshgah Blvd, P.O. 9617976487, Sabzevar, Iran.
Introduction: Low-kV IORT (Low kilovoltage intraoperative radiotherapy) using INTRABEAM machine and dedicated spherical applicators is a candidate modality for breast cancer treatment. The current study aims to quantify the RBE (relative biologic effectiveness) variations of emitted X-rays from the surface of different spherical applicators and bare probe through a hybrid Monte Carlo (MC) simulation approach.
Materials And Methods: A validated MC model of INTRABEAM machine and different applicator diameters, based on GEANT4 Toolkit, was employed for RBE evaluation.
Nanomedicine
November 2020
Department of Radiation Oncology, University of Arkansas for Medical Sciences, Little Rock, AR, United states.
In the present study, Monte Carlo (MC) simulation and analytical calculation methods were used to investigate the potential of cancer treatment for the combination of IORT with nanoparticles (NPs). The Geant4 MC toolkit was used to simulate ZEISS INTRABEAM® IORT machine and its smallest applicator with 1.5 cm diameter.
View Article and Find Full Text PDFZ Med Phys
February 2020
Physics Department, Hakim Sabzevari University, Shohada-e Hastei Blvd, P.O. 9617976487, Sabzevar, Iran.
Low-kV IORT is an increasing modality for breast cancer treatment. Soft X-rays from INTRABEAM (Carl Zeiss Meditec AG, Oberkochen, Germany), a dedicated IORT device along with special spherical applicators are employed for this purpose. A Monte Carlo model of INTRAMBEAM and spherical applicators are introduced in the current study to evaluate the dosimetric and physical characteristics of emitted X-rays from the bare probe and different applicator diameters.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!