Cardiorespiratory monitoring in the delivery room using transcutaneous electromyography.

Arch Dis Child Fetal Neonatal Ed

Department of Neonatology, Emma Children's Hospital, Amsterdam UMC, University of Amsterdam, Amsterdam, North-Holland, Netherlands.

Published: July 2021

Objective: To assess feasibility of transcutaneous electromyography of the diaphragm (dEMG) as a monitoring tool for vital signs and diaphragm activity in the delivery room (DR).

Design: Prospective observational study.

Setting: Delivery room.

Patients: Newborn infants requiring respiratory stabilisation after birth.

Interventions: In addition to pulse oximetry (PO) and ECG, dEMG was measured with skin electrodes for 30 min after birth.

Outcome Measures: We assessed signal quality of dEMG and ECG recording, agreement between heart rate (HR) measured by dEMG and ECG or PO, time between sensor application and first HR read-out and agreement between respiratory rate (RR) measured with dEMG and ECG, compared with airway flow. Furthermore, we analysed peak, tonic and amplitude diaphragmatic activity from the dEMG-based respiratory waveform.

Results: Thirty-three infants (gestational age: 31.7±2.8 weeks, birth weight: 1525±661 g) were included.18%±14% and 22%±21% of dEMG and ECG data showed poor quality, respectively. Monitoring HR with dEMG was fast (median 10 (IQR 10-11) s) and accurate (intraclass correlation coefficient (ICC) 0.92 and 0.82 compared with ECG and PO, respectively). RR monitoring with dEMG showed moderate (ICC 0.49) and ECG low (ICC 0.25) agreement with airway flow. Diaphragm activity started high with a decreasing trend in the first 15 min and subsequent stabilisation.

Conclusion: Monitoring vital signs with dEMG in the DR is feasible and fast. Diaphragm activity can be detected and described with dEMG, making dEMG promising for future DR studies.

Download full-text PDF

Source
http://dx.doi.org/10.1136/archdischild-2020-319535DOI Listing

Publication Analysis

Top Keywords

demg ecg
16
diaphragm activity
12
demg
11
delivery room
8
transcutaneous electromyography
8
vital signs
8
rate measured
8
measured demg
8
airway flow
8
monitoring demg
8

Similar Publications

Introduction: Current cardiorespiratory monitoring in neonates with electrocardiogram (ECG) and chest impedance (CI) has limitations. Adding transcutaneous electromyography of the diaphragm (dEMG) may improve respiratory monitoring, but requires additional hardware. We aimed to determine the feasibility of measuring dEMG and ECG/CI simultaneously using the standard ECG/CI hardware, with its three electrodes repositioned to dEMG electrode locations.

View Article and Find Full Text PDF

Monitoring heart rate (HR) and respiratory rate (RR) is essential in preterm infants and is currently measured with ECG and chest impedance (CI), respectively. However, in current clinical practice these techniques use wired adhesive electrodes which can cause skin damage and hinder parent-infant interaction. Moreover, CI is not always reliable.

View Article and Find Full Text PDF

Cardiorespiratory monitoring in the delivery room using transcutaneous electromyography.

Arch Dis Child Fetal Neonatal Ed

July 2021

Department of Neonatology, Emma Children's Hospital, Amsterdam UMC, University of Amsterdam, Amsterdam, North-Holland, Netherlands.

Objective: To assess feasibility of transcutaneous electromyography of the diaphragm (dEMG) as a monitoring tool for vital signs and diaphragm activity in the delivery room (DR).

Design: Prospective observational study.

Setting: Delivery room.

View Article and Find Full Text PDF

Introduction: Analysis of continuous diaphragm electromyography (dEMG) has not been well studied. We describe a system of analyzing continuous dEMG using implanted electrodes.

Methods: dEMG signal was acquired via two pairs of electrodes near the diaphragm motor points.

View Article and Find Full Text PDF

The selective blockade of potassium channels on excitable membranes by 4-aminopyridine (4-AP) leads to facilitation of neurotransmitter release at a wide variety of synapses. This compound has been shown to be efficacious against lethality induced by saxitoxin (STX) and tetrodotoxin (TTX) in guinea pigs. To characterize the actions of 4-AP in guinea pigs we have investigated its pharmacokinetics (PK) and pharmacodynamics following a 2 mg/kg, intramuscular (im) dose in awake chronically instrumented (IN) animals.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!