Purpose: Treatment plans in proton therapy are more sensitive to uncertainties than in conventional photon therapy. In addition to setup uncertainties, proton therapy is affected by uncertainties in proton range and relative biological effectiveness (RBE). While to date a constant RBE of 1.1 is commonly assumed, the actual RBE is known to increase toward the distal end of the spread-out Bragg peak. Several models for variable RBE predictions exist. We present a framework to evaluate the combined impact and interactions of setup, range, and RBE uncertainties in a comprehensive, variance-based sensitivity analysis (SA).

Material And Methods: The variance-based SA requires a large number (10 -10 ) of RBE-weighted dose (RWD) calculations. Based on a particle therapy extension of the research treatment planning system CERR we implemented a fast, graphics processing unit (GPU) accelerated pencil beam modeling of patient and range shifts. For RBE predictions, two biological models were included: The mechanistic repair-misrepair-fixation (RMF) model and the phenomenological Wedenberg model. All input parameters (patient position, proton range, RBE model parameters) are sampled simultaneously within their assumed probability distributions. Statistical formalisms rank the input parameters according to their influence on the overall uncertainty of RBE-weighted dose-volume histogram (RW-DVH) quantiles and the RWD in every voxel, resulting in relative, normalized sensitivity indices (S = 0: noninfluential input, S = 1: only influential input). Results are visualized as RW-DVHs with error bars and sensitivity maps.

Results And Conclusions: The approach is demonstrated for two representative brain tumor cases and a prostate case. The full SA including RWD calculations took 39, 11, and 55 min, respectively. Range uncertainty was an important contribution to overall uncertainty at the distal end of the target, while the relatively smaller uncertainty inside the target was governed by biological uncertainties. Consequently, the uncertainty of the RW-DVH quantile D for the target was governed by range uncertainty while the uncertainty of the mean target dose was dominated by the biological parameters. The SA framework is a powerful and flexible tool to evaluate uncertainty in RWD distributions and DVH quantiles, taking into account physical and RBE uncertainties and their interactions. The additional information might help to prioritize research efforts to reduce physical and RBE uncertainties and could also have implications for future approaches to biologically robust planning and optimization.

Download full-text PDF

Source
http://dx.doi.org/10.1002/mp.14596DOI Listing

Publication Analysis

Top Keywords

uncertainties proton
12
proton therapy
12
rbe uncertainties
12
rbe
10
uncertainties
9
variance-based sensitivity
8
sensitivity analysis
8
rbe model
8
rbe-weighted dose
8
proton range
8

Similar Publications

Exploration of an adaptive proton therapy strategy using CBCT with the concept of digital twins.

Phys Med Biol

January 2025

Department of Radiology Oncology, Emory University, Clifton Rd, Atlanta, Georgia, 30322-1007, UNITED STATES.

This study aims to develop a digital twin (DT) framework to achieve adaptive proton prostate stereotactic body radiation therapy (SBRT) with fast treatment plan selection and patient-specific clinical target volume (CTV) setup uncertainty. Prostate SBRT has emerged as a leading option for external beam radiotherapy due to its effectiveness and reduced treatment duration. However, interfractional anatomy variations can impact treatment outcomes.

View Article and Find Full Text PDF

Objectives: The purpose of this study was to investigate the fundamental properties of spot-scanning proton beams and compare them to Monte Carlo (MC) simulations, both with and without CT calibration, using spatially diverse combinations of materials.

Methods: A heterogeneous phantom was created by spatially distributing titanium, wax, and thermocol to generate six scenarios of heterogeneous combinations. Proton pencil beams ranging in energy from 100 to 226.

View Article and Find Full Text PDF

An AI dose-influence matrix engine for robust pencil beam scanning protons therapy.

Med Phys

December 2024

National Cancer Center/National Clinical Research Center for Cancer/Cancer Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, People's Republic of China.

Background: Rapid planning is of tremendous value in proton pencil beam scanning (PBS) therapy in overcoming range uncertainty. However, the dose calculation of the dose influence matrix (D) in robust PBS plan optimization is time-consuming and requires substantial acceleration to enhance efficiency.

Purpose: To accelerate the D calculations in PBS therapy, we developed an AI-D engine integrated into our in-house treatment planning system (TPS).

View Article and Find Full Text PDF

Purpose: A novel proton beam delivery method known as DynamicARC spot scanning has been introduced. The current study aims to determine whether the partial proton arc technique, in conjunction with DynamicARC pencil beam scanning (PBS), can meet clinical acceptance criteria for bilateral head and neck cancer (HNC) and provide an alternative to full proton arc and traditional intensity-modulated proton therapy (IMPT).

Method: The study retrospectively included anonymized CT datasets from ten patients with bilateral HNC, all of whom had previously received photon treatment.

View Article and Find Full Text PDF

Proton radiography is a central diagnostic technique for measuring electromagnetic (EM) fields in high-energy-density, laser-produced plasmas. In this technique, protons traverse the plasma where they accumulate small EM deflections which lead to variations in the proton fluence pattern on a detector. Path-integrated EM fields can then be extracted from the fluence image through an inversion process.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!