In Vitro Study of a Stentless Aortic Bioprosthesis Made of Bacterial Cellulose.

Cardiovasc Eng Technol

Cardiac and Vascular Surgery Department, Medical University of Gdańsk, Dębinki 7, 80-211, Gdańsk, Poland.

Published: December 2020

Purpose: The paper present findings from an in vitro experimental study of a stentless human aortic bioprosthesis (HAB) made of bacterial cellulose (BC). Three variants of the basic model were designed and tested to identify the valve prosthesis with the best performance parameters. The modified models were made of BC, and the basic model of pericardium.

Methods: Each model (named V, V and V) was implanted into a 90 mm porcine aorta. Effective Orifice Area (EOA), rapid valve opening time (RVOT) and rapid valve closing time (RVCT) were determined. The flow resistance of each bioprosthesis model during the simulated heart systole, i.e. for the mean differential pressure (ΔP) at the time of full valve opening was measured. All experimental specimens were exposed to a mean blood pressure (MBP) of 90.5 ± 2.3 mmHg.

Results: The V model demonstrated the best performance. The index defining the maximum opening of the bioprosthesis during systole for models V, V and V was 2.67 ± 0.59, 2.04 ± 0.23 and 2.85 ± 0.59 cm, respectively. The mean flow rate through the V valve was 5.7 ± 1, 6.9 ± 0.7 and 8.9 ± 1.4 l/min for stroke volume (SV) of 65, 90 and 110 mL, respectively. The phase of immediate opening and closure for models V, V and V was 8, 7 and 5% of the cycle duration, respectively. The mean flow resistance of the models was: 4.07 ± 2.1, 4.28 ± 2.51 and 5.6 ± 2.32 mmHg.

Conclusions: The V model of the aortic valve prosthesis is the most effective. In vivo tests using BC as a structural material for this model are recommended. The response time of the V model to changed work conditions is comparable to that of a healthy human heart. The model functions as an aortic valve prosthesis in in vitro conditions.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7782396PMC
http://dx.doi.org/10.1007/s13239-020-00500-zDOI Listing

Publication Analysis

Top Keywords

valve prosthesis
12
model
9
study stentless
8
aortic bioprosthesis
8
bacterial cellulose
8
basic model
8
best performance
8
rapid valve
8
valve opening
8
flow resistance
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!