Purpose: To evaluate the flat-panel detector quenching effect and clinical usability of a flat-panel based compact QA device for PBS daily constancy measurements.
Materials & Method: The QA device, named Sphinx Compact, is composed of a 20x20 cm flat-panel imager mounted on a portable frame with removable plastic modules for constancy checks of proton energy (100 MeV, 150 MeV, 200 MeV), Spread-Out-Bragg-Peak (SOBP) profile, and machine output. The potential quenching effect of the flat-panel detector was evaluated. Daily PBS QA tests of X-ray/proton isocenter coincidence, the constancy of proton spot position and sigma as well as the energy of pristine proton beam, and the flatness of SOBP proton beam through the 'transformed' profile were performed and analyzed. Furthermore, the sensitivity of detecting energy changes of pristine proton beam was also evaluated.
Results: The quenching effect was observed at depths near the pristine peak regions. The flat-panel measured range of the distal 80% is within 0.9 mm to the defined ranges of the delivered proton beams. X-ray/proton isocenter coincidence tests demonstrated maximum mismatch of 0.3 mm between the two isocenters. The device can detect 0.1 mm change of spot position and 0.1 MeV energy changes of pristine proton beams. The measured transformed SOBP beam profile through the wedge module rendered as flat.
Conclusions: Even though the flat-panel detector exhibited quenching effect at the Bragg peak region, the proton range can still be accurately measured. The device can fulfill the requirements of the daily QA tests recommended by the AAPM TG224 Report.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.ejmp.2020.10.026 | DOI Listing |
ACS Appl Mater Interfaces
January 2025
Beijing Key Laboratory of Construction-Tailorable Advanced Functional Materials and Green Applications Experimental Center of Advanced Materials, School of Materials Science and Engineering, Beijing Institute of Technology, Beijing 100081, China.
Metal halide perovskites (MHPs) are promising materials for radiation detection. Compared with polycrystalline films, single crystals (SCs) have lower defect density, higher carrier mobility, and lifetime. However, the direct synthesis of MHP SCs for large-area flat panel imaging detectors remains challenging.
View Article and Find Full Text PDFMed Phys
January 2025
Department of Medical Imaging, Radboud University Medical Center, Nijmegen, The Netherlands.
Background: Dedicated breast computed tomography (bCT) systems offer detailed imaging for breast cancer diagnosis and treatment. As new bCT generations are developed, it is important to evaluate their imaging performance and dose efficiency to understand differences over previous models.
Purpose: To characterize the imaging performance and dose efficiency of a second-generation (GEN2) bCT system and compare them to those of a first-generation (GEN1) system.
Int J Part Ther
March 2025
Institute of Medical Physics and Radiation Protection, University of Applied Sciences, Giessen, Germany.
Purpose: The spot size of scanned particle beams is of crucial importance for the correct dose delivery and, therefore, plays a significant role in the quality assurance (QA) of pencil beam scanning ion beam therapy.
Materials And Methods: This study compares 5 detector types-radiochromic film, ionization chamber (IC) array, flat panel detector, multiwire chamber, and IC-for measuring the spot size of proton and carbon ion beams.
Results: Variations of up to 30% were found between detectors, underscoring the impact of detector choice on QA outcomes.
Tuberk Toraks
December 2024
Department of Neurosurgery, Yale University Faculty of Medicine, New Haven, United States.
Introduction: This study aimed to evaluate the imaging findings of the chest flat panel detector computed tomography (FDCT) among coronavirus disease-2019 (COVID-19) positive patients during urgent/emergent interventional neuroradiologic procedures.
Materials And Methods: Chest FDCT examinations were performed using a C-arm mounted FDCT within the interventional radiology (IR) suite if the reverse transcription polymerase chain reaction (RT-PCR) results were pending in patients with clinical findings suggestive of COVID-19. In those who already had positive RT-PCR results, FDCT was performed for acute evaluation only if an acute unexpected cardiopulmonary event occurred during the procedure.
Injury
December 2024
Department of Orthopaedics, Larnaca General Hospital, State Health Services Organisation, Larnaca, Cyprus.
The purpose of this study was to establish typical dose values at orthopaedic operating rooms of the Larnaca General Hospital (LGH). Kerma area product (KAP), fluoroscopy time (FT) and cumulative air-kerma (K) measurements were collected for 821 patients who underwent common and reproducible trauma surgery over a five-year period, with three mobile C-arm systems; two equipped with an image-intensifier and one with a flat-panel detector. Dose indices were automatically extracted from radiation dose structured reports or DICOM meta-data files archived in the PACS, using custom-made software.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!