Characterisation and mapping of scattered radiation fields in interventional radiology theatres.

Sci Rep

CERN, European Organization for Nuclear Research, Geneva, Switzerland.

Published: October 2020

We used the Timepix3 hybrid pixel detector technology in order to determine the exposure of medical personnel to ionizing radiation in an interventional radiology room. We measured the energy spectra of the scattered radiation generated by the patient during X-ray image-guided interventional procedures. We performed measurements at different positions and heights within the theatre. We first observed a difference in fluence for each staff member. As expected, we found that the person closest to the X-ray tube is the most exposed while the least exposed staff member is positioned at the patient's feet. Additionally, we observed a shift in energy from head to toe for practitioners, clearly indicating a non-homogenous energy exposure. The photon counting Timepix3 detector provides a new tool for radiation field characterisation that is easier-to-use and more compact than conventional X-ray spectrometers. The spectral information is particularly valuable for optimising the use of radiation protection gear and improving dosimetry surveillance programs. We also found the device very useful for training purposes to provide awareness and understanding about radiation protection principles among interventional radiology staff.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7599331PMC
http://dx.doi.org/10.1038/s41598-020-75257-5DOI Listing

Publication Analysis

Top Keywords

interventional radiology
12
scattered radiation
8
staff member
8
radiation protection
8
radiation
6
characterisation mapping
4
mapping scattered
4
radiation fields
4
interventional
4
fields interventional
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!