Ring artifacts seriously deteriorate the quality of CT images. Intensity-dependence of detector responses will result in intensity-dependent ring artifacts and time-dependence of CT hardware systems will result in time-dependent ring artifacts. However, only the intensity-dependent ring artifacts are taken into consideration in most post-processing methods. Therefore, the purpose of this study is to propose a general post-processing method, which has a significant removal effect on the intensity-dependent ring artifacts and the time-dependent ring artifacts. First in the proposed method, transform raw CT images into polar coordinate images, and the ring artifacts will manifest as stripe artifacts. Secondly, obtain structure images by smoothing the polar coordinate images and acquire texture images containing some details and stripe artifacts by subtracting the structure images from the polar coordinate images. Third, extract the stripe artifacts from the texture images using mean extraction and texture classification, and obtain the extracted ring artifacts by transforming the extracted stripe artifacts from polar coordinates into Cartesian coordinates. Finally, obtain corrected CT images by subtracting the extracted ring artifacts from the raw CT images, and iterate the corrected CT images in above steps until the ring artifacts extracted in the last iteration are weak enough. Simulation and real data show that the proposed method can remove the intensity-dependent ring artifacts and the time-dependent ring artifacts effectively while preserving image details and spatial resolution. In particular, real data prove that the method is suitable for new CT systems such as the photon counting CT.

Download full-text PDF

Source
http://dx.doi.org/10.1364/OE.401088DOI Listing

Publication Analysis

Top Keywords

ring artifacts
52
artifacts
17
intensity-dependent ring
16
stripe artifacts
16
ring
13
images
13
time-dependent ring
12
polar coordinate
12
coordinate images
12
post-processing method
8

Similar Publications

Ultra-high energy spectral prompt PET.

Phys Med Biol

March 2025

Radiation Detection and Medical Imaging, Delft University of Technology Faculty of Applied Sciences, Mekelweg 15, 2629JB Delft, THE NETHERLANDS, Delft, 2628 CJ, NETHERLANDS.

Objective: Utilizing prompt gammas in preclinical pinhole-collimated PET avoids image degradation due to positron range blurring and photon down scatter, enables multi-isotope PET and can improve counting statistics for low-abundance positron emitters. This was earlier reported for 124I, 89Zr and simultaneous 124I -18F PET using the VECTor scanner (MILabs, The Netherlands), demonstrating sub-mm resolution despite long positron ranges. The aim of the present study is to investigate if such sub-mm PET imaging is also feasible for a large variety of other isotopes including those with extremely high energy prompt gammas (>1 MeV) or with complex emission spectra of prompt gammas.

View Article and Find Full Text PDF

In Computed Tomography (CT) imaging, the ring artifacts caused by the inconsistent detector response can significantly degrade the reconstructed images, having negative impacts on the subsequent applications. The new generation of CT systems based on photon-counting detectors are affected by ring artifacts more severely. The flexibility and variety of detector responses make it difficult to build a well-defined model to characterize the ring artifacts.

View Article and Find Full Text PDF

Background: The supporting structure of high-temperature gas-cooled reactors (HTGR) comprises over 3000 carbon/graphite components, necessitating computed tomography (CT) non-destructive testing before operational deployment as per reactor technical specifications. However, CT images are frequently marred by severe ring artifacts due to the response non-uniformity and non-linearity of detector units, which diminishes the ability to detect defects effectively.

Methods: To address this issue, we propose a physics-based ring artifacts reduction method for CT that employs pixel response correction.

View Article and Find Full Text PDF

DuDo-RAC: Dual-domain optimization for ring artifact correction in photon counting CT.

Comput Methods Programs Biomed

May 2025

Laboratory of Image Science and Technology, School of Computer Science and Engineering, Southeast University, Nanjing, 210096, China; Jiangsu Provincial Joint International Research Laboratory of Medical Information Processing, Southeast University, Nanjing, 210096, China; Key Laboratory of New Generation Artificial Intelligence Technology and Its Interdisciplinary Applications (Southeast University), Ministry of Education, Nanjing, 210096, China.

Background And Objective: Due to the inconsistent response of photon counting detectors (PCDs) pixels to X-rays, there is an obvious presence of low-frequency ring artifacts in CT reconstructed images. Traditional CT ring artifact correction methods are ineffective in correcting low-frequency ring artifacts. Although the pixel-wise polynomial correction method can correct low-frequency ring artifacts, there may still remain residual artifacts due to the inaccuracy in the coefficient measurement and the inability of polynomial functions to perfectly model the relationship between the thickness and post-log raw data.

View Article and Find Full Text PDF

Computed tomography (CT) using synchrotron radiation is a powerful technique that, compared with laboratory CT techniques, boosts high spatial and temporal resolution while also providing access to a range of contrast-formation mechanisms. The acquired projection data are typically processed by a computational pipeline composed of multiple stages. Artifacts introduced during data acquisition can propagate through the pipeline and degrade image quality in the reconstructed images.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!