Objective: To evaluate the effect of slice thickness on CT perfusion analysis of the pancreas in healthy dogs.
Animals: 12 healthy Beagles.
Procedures: After precontrast CT scans, CT perfusion scans of the pancreatic body were performed every second for 30 seconds by sequential CT scanning after injection of contrast medium (iohexol; 300 mg of 1/kg) at a rate of 3 mL/s. Each dog underwent CT perfusion scans twice in a crossover-design study with 2 different slice thicknesses (2.4 and 4.8 mm). Computed tomographic pancreatic perfusion variables, including blood flow, blood volume determined with the maximum slope model, times to the start of enhancement and peak enhancement, permeability, and blood volume determined by Patlak plot analysis, were measured independently by 2 reviewers. The CT perfusion variables were compared between slice thicknesses. Interoperator reproducibility was determined by ICC calculation.
Results: Interoperator reproducibility of CT perfusion variable measurements was excellent on 2.4-mm (mean ± SD ICC, 0.81 ± 0.17) and 4.8-mm (0.90 ± 0.07) slice thicknesses, except for time to peak pancreatic enhancement on 2.4-mm-thick slices, which had moderate reproducibility (intraclass correlation coefficient, 0.473). There was no significant difference in measurements of blood flow, blood volume by either method, times to the start and peak of pancreatic enhancement, or permeability between slice thicknesses.
Conclusions And Clinical Relevance: Results supported that a thin slice thickness of 2.4 mm can be used for assessment of pancreatic perfusion variables in healthy dogs.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.2460/ajvr.81.9.732 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!