Background: A force platform must have validity and reliability for optimal use. The objective of this study was to analyze the validity and the reliability of the Satel 40 Hz stabilometric force platform.
Methods: A study of instrumental validity and reliability, involving a cross-sectional correlational and comparative analysis was performed. To determine the validity, four certified weights located on three axes were used and the ability of the stabilometric force platform to detect changes in the position of the different axes was observed. A test-retest was performed to analyze the reliability. Forty-two symptom-free volunteers participated in the study. Assessments were taken in a standing static position and in a dynamic position, with the eyes open and closed. Three measurements were taken and the intra-class correlation coefficient (ICC) was calculated.
Results: The validity increased as the weight increased for all the variables measured in the stabilometric parameters ( < 0.05). The reliability was shown to be good to excellent for the two visual conditions. The positional variables obtained a higher ICC. The variable with the best ICC was the Y mean in OE (ICC 0.874 and a < 0.001). All the values showed an increase in a dynamic situation.
Conclusion: The findings support the reliability and validity of the Satel 40 Hz stabilometric force platform. The platform could be recommended to evaluate static and dynamic standing balance in healthy adult individuals. Guidelines for treatment and the level of quality of stabilometry could be obtained from its use.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7660083 | PMC |
http://dx.doi.org/10.3390/ijerph17217733 | DOI Listing |
Front Oral Health
January 2025
Oral and Maxillofacial Surgery Department, Faculty of Dental Medicine, Umm Al-Qura University, Makkah, Saudi Arabia.
Background: People experiencing tooth loss need dental prostheses to preserve the integrity of their oral structures and replace the missing teeth. Patient-related outcome measures (PROMs) for perceived barriers to prosthodontics treatment are scarce in the literature.
Aims: The aim of this study was to develop a comprehensive scale to identify and measure barriers to prosthodontic treatment as perceived by partially edentulous patients.
J Int Soc Prev Community Dent
December 2024
Department of Gnathology, San Paolo Dental Building, University of Milan, Milan, Italy.
Aims: To verify the possible association between the stomatognathic and the visual system by analyzing the effects that different mandibular postures may have on the accommodative capacity and ocular convergence and by assessing whether temporomandibular disorders (TMD) can influence visual abilities.
Materials And Methods: A case-control study was performed during a 1-year period on a sample of 100 subjects (50 cases and 50 controls). The subjects underwent a gnathological examination, and two visual tests were performed in three different jaw positions: maximum intercuspation, resting position, and open mouth.
Drug Alcohol Depend Rep
March 2025
Radboud University, Postbus 9102, Nijmegen 6500 HC, the Netherlands.
Introduction: Ecological momentary assessment (EMA) is popular in smoking research to study time-varying processes and design just-in-time personalised cessation interventions. Yet, research examining the psychometric properties of EMA and user experiences with EMA protocols is lacking. We conducted a mixed-methods study to test the EMA component of a mobile intervention for middle to late-aged adolescents (16-20 years) who smoke cigarettes at least weekly.
View Article and Find Full Text PDFBackground: Emergency tracheal intubation is a common and high-risk procedure. Ketamine and etomidate are sedative medicines commonly used to induce anesthesia for emergency tracheal intubation, but whether the induction medication used affects patient outcomes is uncertain.
Research Question: Does the use of ketamine for induction of anesthesia decrease the incidence of death among adults undergoing emergency tracheal intubation, compared to the use of etomidate?
Study Design And Methods: The Randomized trial of Sedative choice for Intubation (RSI) is a pragmatic, multicenter, unblinded, parallel-group, randomized trial being conducted in 14 sites (6 emergency departments and 8 intensive care units) in the United States.
Int J Prev Med
December 2024
Food and Beverages Safety Research Center, Urmia University of Medical Sciences, Urmia, Iran.
Background: An accurate assessment of food intake is necessary to monitor nutritional status. However, differences in cultures and dietary habits between communities make it necessary to create culturally specific tools to evaluate food intake. This study aimed to develop and validate a short food frequency questionnaire (SH-FFQ) in Iranian adults.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!