On the feasibility of absolute 3D dosimetry using LiF thermoluminescence detectors and polymer gels on a 0.35T MR-LINAC.

Phys Med Biol

Department of Medical Physics in Radiation Therapy, German Cancer Research Center, Im Neuenheimer Feld 280, Heidelberg, Germany. National Center for Radiation Research in Oncology (NCRO), Heidelberg Institute for Radiation Oncology (HIRO), Im Neuenheimer Feld 280, Heidelberg, Germany.

Published: October 2020

Background And Purpose: As shown in our previous study, highly accurate absolute dosimetry in 3D is feasible by combining polymer gels (PG) with thermoluminescence dosimetry (TLD). In this setup, the thermoluminescence (TL)-based point dose information is used to renormalize the PG. This new PG-TLD reference system is now extended to measurements in magnetic fields.

Materials And Methods: Experiments were carried out on a conventional 6 MV linear accelerator (LINAC) and a 6 MV 0.35 T magnetic resonance (MR)-LINAC. Signal stability of TLD600 and TLD700 was examined without and with magnetic field. Afterwards, the combination of PAGAT PG and TL detectors was employed within a cylindrical phantom in presence of the magnetic field. Two scenarios were tested: (I) an air-filled phantom and (II) a water-filled phantom. For each scenario, two plans were irradiated: (a) opposed beams with a field size of 10 × 10 cm and (b) a 3D conformal plan assuring homogeneous target coverage using three equally distributed coplanar beams.

Results: Mean relative uncertainty of TL calibration reproducibility for TLD600/TLD700 was 0.49%/0.85% at the MR-LINAC and 0.48%/0.83% for the conventional LINAC. Individual TL calibration coefficients of TLD600 and TLD700 behaved differently in the presence of the magnetic field. An average difference of (3.29 ± 0.89)% occurred for all TLD600, whereas the result for TLD700 is not quite as clear with (1.09 ± 0.89)% after excluding some outliers. Using the TL dose information for PG renormalization, high 3D gamma passing rates were achieved using the 3%/2 mm criteria: 91.0% (Ia), 92.6% (Ib), 94.3% (IIa), 97.4% (IIb).

Conclusion: This study shows that TL signal reproducibility is not affected by a low magnetic field. Nevertheless, absolute calibration coefficients of the individual detectors indicate a dependency on the magnetic field. Hence, a calibration at the appropriate LINAC type is recommended. Furthermore, the previously established renormalization method for PG was applied to measurements at a MR-LINAC and was verified as suitable for evaluations of homogeneous dose distribution in the target volume.

Download full-text PDF

Source
http://dx.doi.org/10.1088/1361-6560/aba6d7DOI Listing

Publication Analysis

Top Keywords

magnetic field
20
absolute dosimetry
8
polymer gels
8
tld600 tld700
8
presence magnetic
8
calibration coefficients
8
magnetic
7
field
6
feasibility absolute
4
dosimetry lif
4

Similar Publications

The electrical conductivity and antibacterial properties are crucial characteristics for bacterial cellulose (BC) based membranes to be broadly applied in the field of wearable electronics. In the study, to achieve these aims, alpha-lipoic acid (LA) was utilized as anchoring groups and reducing agent, hydroxypropyl-β-cyclodextrin (HP-β-CD) capped magnetic particles (FeO NPs) and the in-situ formed silver nanoparticles (AgNPs) were sequentially incorporated into the BC matrix to fabricate BC based nanocomposite membranes (HP-β-CD/FeO/LA@BC and HP-β-CD/FeO/LA/Ag@BC). Fourier transform attenuated total reflectance infrared spectroscopy (FTIR-ATR) and field emission scanning electron microscopy (FE-SEM) analysis proved the dense networks were formed in the modified BC membranes.

View Article and Find Full Text PDF

Purpose: To improve the current method for MRI turbulence quantification which is the intravoxel phase dispersion (IVPD) method. Turbulence is commonly characterized by the Reynolds stress tensor (RST) which describes the velocity covariance matrix. A major source for systematic errors in MRI is the sequence's sensitivity to the variance of the derivatives of velocity, such as the acceleration variance, which can lead to a substantial measurement bias.

View Article and Find Full Text PDF

Magnetic nanoparticles of Nd2Fe14B prepared by ethanol-assisted wet ball milling technique.

Sci Rep

January 2025

Environmental and Occupational Hazards Control Research Center, Research Institute for Health Sciences and Environment, Shahid Beheshti University of Medical Sciences, Tehran, Iran.

The magnetic material Nd2Fe14B is one of the strongest magnetic materials found in nature. The demand for the production of these nanoparticles is significantly high due to their exceptional properties. The aim of the present study is to synthesize magnetic nanoparticles of Nd2Fe14B using ethanol in the wet ball milling technique (WBMT).

View Article and Find Full Text PDF

Carbon-supported Fe single atom nanozymes with long-lasting ROS generation and high NIR photothermal performance for synergistic cancer therapy.

J Colloid Interface Sci

April 2025

High Magnetic Field Laboratory, Hefei Institutes of Physical Science, Chinese Academy of Science, Hefei, Anhui 230031, PR China; University of Science and Technology of China, Hefei, Anhui 230026, PR China. Electronic address:

Synergistic therapy combining photothermal therapy (PTT) and chemodynamic therapy (CDT) has proven to be a highly effective strategy for cancer treatment. However, PTT heavily relies on the accumulation of therapeutic agents at the tumor site. The peroxidase (POD) activity of common catalysts can be rapidly exhausted during the accumulation process, prior to laser intervention, thereby diminishing the synergistic enhancement effect of the combined therapy.

View Article and Find Full Text PDF

Overactive bladder (OAB) is a prevalent chronic condition affecting approximately 12% of adults, with incidence increasing with age. While pharmacological and behavioural therapies are standard treatments, their efficacy is often limited by side effects and poor adherence. This study aimed to compare the therapeutic effects of precision magnetic stimulation guided by motor-evoked potential with general magnetic therapy in patients with OAB.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!