Purpose: With abdominal magnetic resonance elastography (MRE) often suffering from breathing artifacts, it is recommended to perform MRE during breath-hold. However, breath-hold acquisition prohibits extended multifrequency MRE examinations and yields inconsistent results when patients cannot hold their breath. The purpose of this work was to analyze free-breathing strategies in multifrequency MRE of abdominal organs.
Methods: Abdominal MRE with 30, 40, 50, and 60 Hz vibration frequencies and single-shot, multislice, full wave-field acquisition was performed four times in 11 healthy volunteers: once with multiple breath-holds and three times during free breathing with ungated, gated, and navigated slice adjustment. Shear wave speed maps were generated by tomoelastography inversion. Image registration was applied for correction of intrascan misregistration of image slices. Sharpness of features was quantified by the variance of the Laplacian.
Results: Total scan times ranged from 120 seconds for ungated free-breathing MRE to 376 seconds for breath-hold examinations. As expected, free-breathing MRE resulted in larger organ displacements (liver, 4.7 ± 1.5 mm; kidneys, 2.4 ± 2.2 mm; spleen, 3.1 ± 2.4 mm; pancreas, 3.4 ± 1.4 mm) than breath-hold MRE (liver, 0.7 ± 0.2 mm; kidneys, 0.4 ± 0.2 mm; spleen, 0.5 ± 0.2 mm; pancreas, 0.7 ± 0.5 mm). Nonetheless, breathing-related displacement did not affect mean shear wave speed, which was consistent across all protocols (liver, 1.43 ± 0.07 m/s; kidneys, 2.35 ± 0.21 m/s; spleen, 2.02 ± 0.15 m/s; pancreas, 1.39 ± 0.15 m/s). Image registration before inversion improved the quality of free-breathing examinations, yielding no differences in image sharpness to uncorrected breath-hold MRE in most organs (P > .05).
Conclusion: Overall, multifrequency MRE is robust to breathing when considering whole-organ values. Respiration-related blurring can readily be corrected using image registration. Consequently, ungated free-breathing MRE combined with image registration is recommended for multifrequency MRE of abdominal organs.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1002/mrm.28558 | DOI Listing |
Sci Rep
November 2024
Department of Radiology, Charité-Universitätsmedizin Berlin, Corporate Member of Freie Universität Berlin, Humboldt-Universität zu Berlin, Berlin Institute of Health, Charitéplatz 1, 10117, Berlin, Germany.
Magnetic resonance elastography (MRE) is a noninvasive brain stiffness mapping method. Ultrasound-based transtemporal time-harmonic elastography (THE) is emerging as a cost-effective, fast alternative that has potential applications for bedside monitoring of intracranial pressure. We aim to investigate the accuracy of THE in comparison to MRE performed in the brain.
View Article and Find Full Text PDFJ Gastrointest Surg
November 2024
Department of Pancreatobiliary Surgery, The First Affiliated Hospital, Sun Yat-sen University, Guangzhou, Guangdong, China. Electronic address:
Background: Clinically relevant postoperative pancreatic fistula (CR-POPF) is the major complication of pancreatoduodenectomy, and the pancreatic texture is one of the potential affecting factors. Multifrequency magnetic resonance elastography (MRE) is a novel technique for measuring tissue stiffness, but its value in predicting CR-POPF preoperatively has not been well documented.
Methods: A total of 70 patients who underwent multifrequency MRE before pancreatoduodenectomy between July 2021 and April 2024 were retrospectively recruited into the study.
J Magn Reson Imaging
October 2024
Department of Radiology, University Medical Center Groningen, Groningen, The Netherlands.
Background: Cardiac magnetic resonance elastography (MRE) shows promise in assessing the mechanofunctional properties of the heart but faces clinical challenges, mainly synchronization with cardiac cycle, breathing, and external harmonic stimulation.
Purpose: To determine the reproducibility of in vivo cardiac multifrequency MRE (MMRE) for assessing diastolic left ventricular (LV) stiffness and viscosity.
Study Type: Prospective.
Front Oncol
September 2024
Department of Neurosurgery, Charité - Universitätsmedizin Berlin, Corporate Member of Freie Universität Berlin and Humboldt Universität zu Berlin, Berlin, Germany.
Introduction: Magnetic Resonance Elastography (MRE) allows the non-invasive quantification of tumor biomechanical properties . With increasing incidence of brain metastases, there is a notable absence of appropriate preclinical models to investigate their biomechanical characteristics. Therefore, the purpose of this work was to assess the biomechanical characteristics of B16 melanoma brain metastases (MBM) and compare it to murine GL261 glioblastoma (GBM) model using multifrequency MRE with tomoelastography post processing.
View Article and Find Full Text PDFJ Mech Behav Biomed Mater
December 2024
Charité - Universitätsmedizin Berlin, corporate member of Freie Universität Berlin and Humboldt-Universität zu Berlin, Department of Radiology, Hindenburgdamm 30, 12203 Berlin, Germany.
Purpose: To study the potential of viscoelastic parameters such as liver stiffness, loss tangent (marker of viscous properties) and viscoelastic dispersion to detect hepatic inflammation by in-vivo and ex-vivo MR elastography (MRE) at low and high vibration frequencies.
Methods: 15 patients scheduled for liver tumor resection surgery were prospectively enrolled in this IRB-approved study and underwent multifrequency in-vivo MRE (30-60Hz) at 1.5-T prior to surgery.