Background: A triple-coaxial (triaxial) system, which consists of a small microcatheter, a large microcatheter, and a 4-Fr. catheter, has been reported to allow super-selective catheterization. A 1.6-Fr. microcatheter has recently become available as the small microcatheter for the triaxial system, in addition to 0.014-in. pushable bare platinum coils that may be introduced into the 1.6-Fr. microcatheter.

Purpose: The purpose of the present study was to evaluate the feasibility of 0.014-in. pushable bare platinum coils in embolization through the 1.6-Fr. microcatheter of the triaxial system.

Material And Methods: Between November 2015 and October 2019, 19 embolizations were performed on 18 patients, 9 males and 9 females with a median age of 77 years (range, 41-88 years), using 0.014-in. pushable bare platinum coils through the 1.6-Fr. microcatheter of the triaxial system. The technical success rate, clinical success rate, and complications associated with the procedure were assessed. Technical success was defined as the successful delivery and placement of 0.014-in. pushable bare platinum coils, and clinical success as the immediate postembolic complete cessation of blood flow confirmed by digital subtraction angiography.

Results: Eighty-four 0.014-in. pushable bare platinum coils were delivered and 19 arteries were successfully embolized. The median number of 0.014-in. pushable bare platinum coils was 4 (range, 1-12). The technical success rate was 100% (84/84) and the clinical success rate was also 100% (19/19). There were no complications associated with the procedures.

Conclusion: The use of 0.014-in. pushable bare platinum coils in super-selective embolization through the 1.6-Fr. microcatheter of the triaxial system appears to be feasible and safe.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7549165PMC
http://dx.doi.org/10.1177/2058460120964066DOI Listing

Publication Analysis

Top Keywords

0014-in pushable
32
pushable bare
32
bare platinum
32
platinum coils
32
16-fr microcatheter
20
microcatheter triaxial
20
triaxial system
20
success rate
16
embolization 16-fr
12
technical success
12

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!