Evaluate the effect of corneal thickness, densitometry and curvature on intraocular pressure (IOP) measurements obtained by Goldmann applanation tonometry (GAT), non-contact tonometry (NCT), rebound tonometry (RT), and dynamic contour tonometry (DCT). A cross-sectional prospective study involving 40 participants was performed. Corneal measurements were obtained using Pentacam (Oculus GMbH, Wetzlar, Germany), densitometry was measured at annuli of 0-2, 2-6, 6-10 and 10-12 mm. The relationship between corneal thickness (central, 4 and 6 mm), corneal astigmatism and corneal densitometry and IOP was examined. There was a significant relationship between corneal thickness (central, 4 and 6 mm) and GAT180, GAT90, RT, and NCT ( < 0.001 for all comparisons) but not for DCT. Higher corneal densitometry (6-10 mm and 10-12 mm zones) was associated with higher IOP from GAT180 and GAT90, and higher densitometry in the 6-10 mm zone correlated with higher IOP from NCT, however corneal densitometry increased with age. Accounting for age, the relationship between corneal densitometry and IOP measurements was not significant. In eyes with greater corneal astigmatism there was a greater difference between GAT90 and GAT180 measurements. IOP measurements may be affected by corneal thickness, densitometry and curvature. DCT was less affected by properties of the cornea compared to other devices.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7711663PMC
http://dx.doi.org/10.3390/vision4040045DOI Listing

Publication Analysis

Top Keywords

corneal thickness
20
corneal densitometry
16
corneal
12
thickness densitometry
12
densitometry curvature
12
iop measurements
12
relationship corneal
12
densitometry
9
curvature intraocular
8
intraocular pressure
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!