Biomechanical evaluation of three different configurations of external fixators for treating distal third tibia fracture: Finite element analysis in axial, bending and torsion load.

Comput Biol Med

Medical Devices and Technology Centre (MEDiTEC), Institute of Human Centered Engineering (iHumEn), Universiti Teknologi Malaysia, 81310, Skudai, Johor, Malaysia; Bioinspired Devices and Tissue Engineering (BIOINSPIRA) Group, School of Biomedical Engineering & Health Sciences, Faculty of Engineering, Universiti Teknologi Malaysia, 81310, Skudai, Johor, Malaysia. Electronic address:

Published: December 2020

External fixators have been widely used in treating open fractures and have produced excellent outcomes, as they could successfully heal bones. The stability of external fixators lies greatly in their construction. Factors that associated with the stability of the external fixators includes stress, displacement, and relative micromotion. Three-dimensional (3D) models of bone and external fixators were constructed by using 3D modelling software, namely Materialise and SolidWorks, respectively. Three different configurations of external fixators namely Model 1, Model 2, and Model 3 were analysed. Three load cases were simulated to assess the abovementioned factors at the bone, specifically at the fracture site and at the external fixator. Findings showed that the double-cross configuration (Model 3) was the most promising in axial, bending, and torsion load cases as compared to the other two configurations. The no-cross configuration (Model 1) had the highest risk of complication due to high stress, relative micromotion, and displacement in the bending and torsion load cases. On the other hand, the single-cross configuration (Model 2) had the highest risk of complication when applied with axial load. In conclusion, the double-cross locking construct (Model 3) showed the biggest potential to be a new option for medical surgeons in treating patients associated with bone fracture. This new double-cross locking construct showed superior biomechanical stability as compared to single-cross and no-cross configurations in the axial, bending, and torsion load cases.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.compbiomed.2020.104062DOI Listing

Publication Analysis

Top Keywords

external fixators
24
bending torsion
16
torsion load
16
load cases
16
axial bending
12
configuration model
12
three configurations
8
configurations external
8
fixators treating
8
stability external
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!