Objective: In this study, we aimed to evaluate the efficiency of rotational thrombectomy device in pharmaco-mechanical thrombolysis for symptomatic acute ilio-femoral deep vein thrombosis.
Materials And Methods: Between August 2013 and May 2018,82 patients with acute deep vein thrombosis comprising the iliofemoral segment whom underwent Pharmaco-mechanical thrombolysis were evaluated retrospectively. The Cleaner thrombectomy device was used. The resolution of thrombi was examined and graded. Development of post-thrombotic syndrome was assessed with Villalta scores.
Results: 75 patients (91.4%) had complete thrombus resolution. Between 50-99% resolution was noted in 6 patients (7.4%) and in one (1.2%) case less than %50 thrombus resolution was obtained. Seventy-five patients (91%) of the cohort could be treated in a single session; 7 patients (8.6%) required reintervention(s). Although improved post-thrombotic syndrome rates were lower at the short term, Villalta scores gradually increased during follow up.
Conclusions: In conclusion, Pharmaco-mechanical thrombolysis with Cleaner thrombectomy device is a safe and beneficial method for the treatment of acute iliofemoral deep vein thrombosis. Long term follow up data of large volume multicenter studies are warranted.