Aims: To establish a simple criterion for determining a failed His-bundle pacing (HBP). This criterion states that if stimulus to QRS end interval is longer than His-bundle potential to QRS end interval ('S-QRSend > H-QRSend') then a failed HBP can be determined.

Methods And Results: We performed retrospective analysis on 737 pacing tests around His-bundle in 241 patients and prospective analysis on 400 tests in 123 patients. A successful HBP is defined as that whole His-bundle is captured with or without capture of adjacent ventricular myocardium, otherwise, a failed HBP was considered. The output criteria and effective refractory period criteria were used as the gold standards for determining a successful HBP. The gold standards are that if decreasing the pacing output or pacing cycle length to a certain level results in duration or morphology changes of QRS, then a successful HBP is ascertained. In retrospective analysis of patients with normal His-Purkinje conduction, a failed HBP was determined in 31% (154/492) of pacing tests according to 'S-QRSend > H-QRSend'; all of them were validated by the gold standards (specificity = 100%). In prospective study, a failed HBP was confirmed according to the simple criterion with 100% accuracy in 33% (79/241) pacing tests. This simple criterion was also suitable for patients with His-Purkinje conduction disease although cases with 'S-QRSend > H-QRSend' rarely occurred.

Conclusion: A failed HBP can be easily and reliably determined solely by 'S-QRSend > H-QRSend' in more than 30% pacing tests.

Download full-text PDF

Source
http://dx.doi.org/10.1093/europace/euaa244DOI Listing

Publication Analysis

Top Keywords

failed hbp
20
's-qrsend h-qrsend'
16
pacing tests
16
simple criterion
12
successful hbp
12
gold standards
12
hbp
9
criterion determining
8
determining failed
8
failed his-bundle
8

Similar Publications

His bundle pacing combined with atrioventricular node ablation for atrial fibrillation: a systematic review and meta-analysis.

Expert Rev Med Devices

November 2024

Department of Cardiology, Laboratory of Heart Center, Zhujiang Hospital, Southern Medical University, Guangzhou, Guangdong, China.

Introduction And Objective: His bundle pacing (HBP) could replace failed biventricular pacing (BVP) in guidelines (IIa Indication), but the high capture thresholds and backup lead pacing requirements limit its development. We assessed the efficacy and safety of HBP combined with atrioventricular node ablation (AVNA) for atrial fibrillation (AF) and compared with BVP and left bundle branch pacing (LBBP).

Methods: We reviewed PubMed, Embase, Web of Science, and Cochrane Library databases on left ventricular ejection fraction (LVEF), New York Heart Association (NYHA) score, QRS duration (QRSd), and pacing threshold.

View Article and Find Full Text PDF

Left ventricular volumes and function in successful and failed His-BundLe Pacing. A comparative prospective study.

J Cardiovasc Electrophysiol

November 2024

Department of Medicine and Surgery, Varese and Department of Medicine and Cardiopulmonary Rehabilitation, Maugeri Care and Research Institute, IRCCS Tradate, University of Insubria, Varese, Italy.

Introduction: Initial data suggest that His Bundle Pacing (HBP) could preserve long-term cardiac structure and function better than Right Ventricular Pacing (RVP), but evidence is limited.

Methods: We studied consecutive patients with baseline ejection fraction (EF) ≥ 50% who underwent HBP attempt, either successful (HBP group) or failed (RVP group). Two-dimensional (2D) and three-dimensional (3D) echocardiography were carried out at baseline and after 6 months of ventricular pacing burden > 20%.

View Article and Find Full Text PDF

The current gold standard in device therapy for advanced heart failure (HF), which has been firmly established in HF management for more than 25 years, is classical biventricular pacing (BiV-CRT). In the last decade, a new pacing modality called conduction system pacing (CSP) has emerged as a variant for advanced cardiac device therapy. It provides pacing with preserved intrinsic cardiac activation by direct stimulation of the specific cardiac conduction system.

View Article and Find Full Text PDF

Objective: To define the concept of surgeon-scientists and identify the root causes of their decline in number and impact. The secondary aim was to provide actionable remedies.

Background: Surgeons who conduct research in addition to patient care are referred to as "surgeon-scientists.

View Article and Find Full Text PDF

Background: Experience with the transvenous extraction of leads used for His bundle pacing (HBP) is limited.

Methods: Analysis of 3897 extractions including 27 HBP and 253 LVP (left ventricular pacing) leads.

Results: The main reason for HBP lead extraction was lead failure (59.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!