Aim: To evaluate the effectiveness of Grand Seiko Ref/Keratometer WAM-5500 compared to Topcon KR800 autorefractor in detecting refractive error in large scale vision screening for Chinese school age children with the WHO criteria.

Methods: A total of 886 participants were enrolled with mean age of 9.49±1.88y from Tianjin, China. Spherical equivalent (SE) was obtained from un-cycloplegic autorefraction and cycloplegic autorefraction. Topcon KR 800 (Topcon) and Grand Seiko WAM-5500 (WAM) autorefractors were used. Bland-Altman Plot and regression were generated to compare their performance. The overall effectiveness of detecting early stage refractive error was analyzed with receiver operating characteristic (ROC) curves.

Results: The mean SE was -0.98±1.81 diopter (D) and the prevalence of myopia was 48.9% defined by WHO criteria according to the result of cycloplegic autorefraction. The mean SE of un-cycloplegic autorefraction with Topcon and WAM were -1.21±1.65 and -1.20±1.68 D respectively. There was a strong linear agreement between result obtained from WAM and cycloplegic autorefraction with an of 0.8318. Bland-Altman plot indicated a moderate agreement of cylinder values between the two methods. The sensitivity and specificity for detecting hyperopia were 90.52% and 83.51%; for detecting myopia were 95.60% and 90.24%; for detecting astigmatism were 79.40% and 90.21%; for detecting high myopia were 98.16% and 98.91% respectively.

Conclusion: These findings suggest that both Grand Seiko and Topcon autorefractor can be used in large-scale vision screening for detecting refractive error in Chinese population. Grand Seiko gives relatively better performance in detecting myopia, hyperopia, and high myopia for school age children.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7511371PMC
http://dx.doi.org/10.18240/ijo.2020.10.22DOI Listing

Publication Analysis

Top Keywords

grand seiko
16
vision screening
12
school age
12
age children
12
refractive error
12
cycloplegic autorefraction
12
large scale
8
scale vision
8
screening chinese
8
chinese school
8

Similar Publications

The peripheral retinal refractive state plays an important role in eye growth and development and is closely related to the development of myopia. Existing methods for measuring the peripheral retinal refractive state are cumbersome and can only detect in a limited range. To address the above shortcomings, this paper proposes a retinal refractive state detection method using optical refractive compensation imaging.

View Article and Find Full Text PDF

Peripheral Eye Length Evaluation in Myopic Children Undergoing Orthokeratology Treatment for 12 Months Using MRI.

Clin Optom (Auckl)

February 2024

Department of Electrical, Electronics and Systems, Faculty of Engineering and Built, UKM, Bangi, Selangor, Malaysia.

Purpose: To investigate changes in peripheral eye length (PEL) in myopic children undergoing orthokeratology (Ortho-K) treatment for 12 months using MRI. The results were compared to single vision spectacle wearers (SVS).

Patients And Methods: A total of 70 children with myopia (aged 8-9 years old) were recruited.

View Article and Find Full Text PDF

Purpose: This study compares the trends of changes in corneal asphericity, corneal and total aberrations and peripheral refraction in myopic eyes after excimer laser and orthokeratology correction.

Material And Methods: Aberrometry (OPD-Scan III; Nidek, Japan) was performed in 63 patients (126 eyes) with moderate and high myopia before and after femtosecond laser-assisted in situ keratomileusis (Femto-LASIK; 88 eyes, group 1) and photorefractive keratectomy (PRK; 38 eyes, group 2). Peripheral refraction (Grand Seiko AutoRef/Keratometer) at 15° and 30° from the center of the fovea was observed in 12 patients of group 1 and in 18 patients with a background of orthokeratology correction (group 3).

View Article and Find Full Text PDF

Introduction: This study evaluated novel automatic dual rotational Risley prisms (ADRRPs) as a vergence exercise tool for patients with myopia to improve accommodative lag and accommodative facility.

Methods: Participants with myopia aged 20-24 years were recruited. After vergence exercises with prisms (treatment group) or plano lenses (control group) using ADRRPs for 10 min, measurements were taken using an open-field autorefractor (Grand Seiko WAM-5500) at viewing distances of 0.

View Article and Find Full Text PDF

Pupillary dynamics and accommodative response in mild traumatic brain injury.

Taiwan J Ophthalmol

May 2023

Department of Binocular Vision and Neuro-Optometry, Sankara Nethralaya, Chennai, Tamil Nadu, India.

Purpose: To measure the pupillary dynamics and accommodative response in individuals with mild traumatic brain injury (mTBI) as compared to age-matched controls.

Materials And Methods: This prospective comparative study was carried out at the neuro-optometry clinic of a tertiary eye care hospital. Sixty-three subjects with a history of mTBI and ninety age-matched controls were enrolled in this study.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!