This study reports a novel impedimetric immunosensor for protein D detection in purified and bacterial (Haemophilus influenzae, Hi) samples. The detection was based on antigen recognition by anti-protein D antibodies (apD) immobilised at the maze-like boron-doped carbon nanowall electrodes (B:CNW). The B:CNW electrodes were synthesised, and their surface was characterised by scanning electron microscopy (SEM) and X-ray photoelectron spectroscopy (XPS) methods. The sensor was prepared in a two-step procedure: apD were covalently linked on the previously modified B:CNW electrodes using diazonium salt. Modification steps were controlled by electrochemical impedance spectroscopy (EIS) and cyclic voltammetry (CV) measurements. The immunosensor exhibited excellent electrochemical performance, stability, satisfactory sensitivities, and linear ranges for antigen detection. Protein D was detected down to 2.39 × 10fg/mL with a linear range extending from 3.37 × 10to 3.37 × 10μg/mL (in purified sample). Next, Hi's LOD was 5.20 × 10CFU/mL with a linear range of 8.39 × 10-8.39 × 10CFU/mL. Selectivity studies showed no reaction with negative samples as Streptococcus pyogenes, Streptococcus pneumoniae or Bordetella parapertussis bacteria. Therefore, the new approach is suitable for rapid and quantitative detection of Hi, and is a good candidate for further tests on clinical samples.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.talanta.2020.121623DOI Listing

Publication Analysis

Top Keywords

haemophilus influenzae
8
maze-like boron-doped
8
boron-doped carbon
8
carbon nanowall
8
nanowall electrodes
8
bcnw electrodes
8
linear range
8
highly selective
4
selective impedimetric
4
impedimetric determination
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!