Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Purpose: Accelerated partial breast irradiation (APBI) is alternative treatment option for patients with early stage breast cancer. The interplay effect on volumetric modulated arc therapy APBI (VMAT-APBI) has not been clarified. This study aimed to evaluate the feasibility of VMAT-APBI for patients with small breasts and investigate the amplitude of respiratory motion during VMAT-APBI delivery that significantly affects dose distribution.
Methods: The VMAT-APBI plans were generated with 28.5 Gy in five fractions. We performed patient-specific quality assurance using Delta phantom under static conditions. We also measured point dose and dose distribution using the ionization chamber and radiochromic film under static and moving conditions of 2, 3 and 5 mm. We compared the measured and calculated point doses and dose distributions by dose difference and gamma passing rates.
Results: A total of 20 plans were generated; the dose distributions were consistent with those of previous reports. For all measurements under static conditions, the measured and calculated point doses and dose distributions showed good agreement. The dose differences for chamber measurement were within 3%, regardless of moving conditions. The mean gamma passing rates with 3%/2 mm criteria in the film measurement under static conditions and with 2 mm, 3 mm, and 5 mm of amplitude were 95.0 ± 2.0%, 93.3 ± 3.3%, 92.1 ± 6.2% and 84.8 ± 7.8%, respectively. The difference between 5 mm amplitude and other conditions was statistically significant.
Conclusions: Respiratory management should be considered for the risk of unintended dose distribution if the respiratory amplitude is >5 mm.