Investigating the use of virtual 4DCT from 4DMRI in gated carbon ion radiation therapy of abdominal tumors.

Z Med Phys

Department of Electronics, Information and Bioengineering, Politecnico di Milano, Milano 20133, Italy; Centro Nazionale di Adroterapia Oncologica, Pavia 27100, Italy.

Published: February 2022

Purpose: To generate virtual 4DCT from 4DMRI with field of view (FOV) extended to the entire involved patient anatomy, in order to evaluate its use in carbon ion radiation therapy (CIRT) of the abdominal site in a clinical scenario.

Materials And Methods: The virtual 4DCT was generated by deforming a reference CT in order to (1) match the anatomy depicted in the 4DMRI within its FOV, by calculating deformation fields with deformable image registration to describe inter-fractional and breathing motion, and (2) obtain physically plausible deformation outside of the 4DMRI FOV, by propagating and modulating the previously obtained deformation fields. The implemented method was validated on a digital anthropomorphic phantom, for which a ground truth (GT) 4DCT was available. A CIRT treatment plan was optimized at the end-exhale reference CT and the RBE-weighted dose distribution was recalculated on both the virtual and GT 4DCTs. The method estimation error was quantified by comparing the virtual and GT 4DCTs and the corresponding recomputed doses. The method was then evaluated on 8 patients with pancreas or liver tumors treated with CIRT using respiratory gating at end-exhale. The clinical treatment plans adopted at the National Center for Oncological Hadrontherapy (CNAO, Pavia, Italy) were considered and the dose distribution was recomputed on all respiratory phases of the planning and virtual 4DCTs. By comparing the two datasets and the corresponding dose distributions, the geometrical and dosimetric impact of organ motion was assessed.

Results: For the phantom, the error outside of the 4DMRI FOV was up to 4.5mm, but it remained sub-millimetric in correspondence to the target within the 4DMRI FOV. Although the impact of motion on the target D resulted in variations ranging from 22% to 90% between the planned dose and the doses recomputed on the GT 4DCT phases, the corresponding estimation error was ≤2.2%. In the patient cases, the variation of the baseline tumor position between the planning and the virtual end-exhale CTs presented a median (interquartile range) value of 6.0 (4.9) mm. For baseline variations larger than 5mm, the tumor D variation between the plan and the dose recomputed on the end-exhale virtual CT resulted larger than 10%. Median variations higher than 10% in the target D and gastro-intestinal OARs D were quantified at the end-inhale, whereas close to the end-exhale phase, limited variations of relevant dose metrics were found for both tumor and OARs.

Conclusions: The negligible impact of the geometrical inaccuracy in the estimated anatomy outside of the 4DMRI FOV on the overall dosimetric accuracy suggests the feasibility of virtual 4DCT with extended FOV in CIRT of the abdominal site. In the analyzed patient group, inter-fractional variations such as baseline variation and breathing variability were quantified, demonstrating the method capability to support treatment planning in gated CIRT of the abdominal site.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9948849PMC
http://dx.doi.org/10.1016/j.zemedi.2020.08.005DOI Listing

Publication Analysis

Top Keywords

4dmri fov
20
virtual 4dct
16
cirt abdominal
12
abdominal site
12
virtual 4dcts
12
4dct 4dmri
8
carbon ion
8
ion radiation
8
radiation therapy
8
virtual
8

Similar Publications

Investigating the use of virtual 4DCT from 4DMRI in gated carbon ion radiation therapy of abdominal tumors.

Z Med Phys

February 2022

Department of Electronics, Information and Bioengineering, Politecnico di Milano, Milano 20133, Italy; Centro Nazionale di Adroterapia Oncologica, Pavia 27100, Italy.

Purpose: To generate virtual 4DCT from 4DMRI with field of view (FOV) extended to the entire involved patient anatomy, in order to evaluate its use in carbon ion radiation therapy (CIRT) of the abdominal site in a clinical scenario.

Materials And Methods: The virtual 4DCT was generated by deforming a reference CT in order to (1) match the anatomy depicted in the 4DMRI within its FOV, by calculating deformation fields with deformable image registration to describe inter-fractional and breathing motion, and (2) obtain physically plausible deformation outside of the 4DMRI FOV, by propagating and modulating the previously obtained deformation fields. The implemented method was validated on a digital anthropomorphic phantom, for which a ground truth (GT) 4DCT was available.

View Article and Find Full Text PDF

Optimizing 4D abdominal MRI: image denoising using an iterative back-projection approach.

Phys Med Biol

January 2020

'Institut de Mathématiques de Bordeaux', University of Bordeaux/CNRS UMR 5251, 351 Cours de la Libération, 33405 Talence Cedex, France. Author to whom any correspondence should be addressed.

4D-MRI is a promising tool for organ exploration, target delineation and treatment planning. Intra-scan motion artifacts may be greatly reduced by increasing the imaging frame rate. However, poor signal-to-noise ratios (SNR) are observed when increasing spatial and/or frame number per physiological cycle, in particular in the abdomen.

View Article and Find Full Text PDF

Purpose: 4D-MRI based on body area (BA) surrogate using axial image acquisition has been demonstrated. Since respiratory motion mostly occurs in the superior-inferior (SI) direction, it is expected that sagittal acquisition may provide more robust and accurate breathing signal than axial acquisition. The aim of this study was to investigate the feasibility of extracting breathing signals from sagittal images using BA surrogate and its application in 4D-MRI.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!