Objective: The aim of this study was to determine the effect of insertion torque and angulation on the push-out strength of screws in Atraumatic Rigid Fixation (ARIX) system.
Materials And Methods: mechanical tests of the ARIX system were conducted. Screw plate constructs ( = 120) were tested using five different insertion torques at four different angles relative to the perpendicular axis of the plate. Before the push-out test, screws were locked into the plates, and the push-out force of the screw was measured by applying a load parallel to the screw axis.
Results: Implant failure was observed at 0.8 Nm at an insertion angle of 15 degrees, and at 1 Nm at 0 degree, 5 degrees and 10 degrees. Two types of failures were observed: cold welding and plate deformation. An insertion torque of 0.8 Nm produced a significantly higher push-out force compared with 0.2 Nm. Non-angled specimens with 0.8 Nm insertion torque exhibited significantly higher screw push-out forces compared with other tested specimens and insertion angles. Insertion angle did not affect screw push-out force at insertion torques of 0, 0.2, 0.4 and 0.6 Nm.
Conclusion: The ARIX locking system is much more sensitive to insertion torque than angle. An effect of insertion angle was observed only at an insertion torque of 0.8 Nm, under which all angles significantly decreased push-out force relative to zero degrees of angulation. In addition, low insertion torques can result in screws loosening over time, while greater insertion torques than 1 Nm can result in screw head stripping and plate hole deformation.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1055/s-0040-1716679 | DOI Listing |
Bioengineering (Basel)
January 2025
Department of Medical, Oral and Biotechnological Sciences, "G. D'Annunzio" University of Chieti-Pescara, 66100 Chieti, Italy.
This study aimed to evaluate the biomechanical performance of two truncated cone implant designs in maxillary sinus lift (MSL) procedures using polyurethane laminas. A total of 128 implants were used. Polyurethane laminas were divided into two groups based on thickness (1 and 3 mm) and two subgroups based on density (20 and 30 pounds per cubic foot, PCF).
View Article and Find Full Text PDFVet Comp Orthop Traumatol
January 2025
School of Engineering, University of Guelph, Guelph, Ontario, Canada.
Objective: To determine the effect of locking head inserts (LHI) on plate strain, stiffness, and deformation when applied to a 3.5-mm broad locking compression plate (LCP) in an open fracture-gap model.
Study Design: Six, 13-hole, 3.
Mater Horiz
January 2025
School of Materials Science and Engineering, Key Laboratory of Advanced Materials and Devices for Post-Moore Chips, Ministry of Education, University of Science and Technology Beijing, Beijing 100083, China.
Realizing spin-orbit torque (SOT)-driven magnetization switching offers promising opportunities for the advancement of next-generation spintronics. However, the relatively low charge-spin conversion efficiency accompanied by an ultrahigh critical switching current density () remains a significant obstacle to the further development of SOT-based storage elements. Herein, spin absorption engineering at the ferromagnet/nonmagnet interface is firstly proposed to achieve high SOT efficiency in Pt/Co/Ir trilayers.
View Article and Find Full Text PDFNatl J Maxillofac Surg
November 2024
Department of Prosthodontics, BBD CODS, Lucknow, Uttar Pradesh, India.
Introduction: Dental implants are the most popular alternative to rehabilitation of missing teeth and oro-facial structures. The outcome of such procedures depends on various factors and most-importantly on the osseo-integation with the surrounding bone. The evaluation needs better visualization and evaluation using CBCT anaylsis and determination of HU, using an appropriate software.
View Article and Find Full Text PDFTissue Eng Regen Med
January 2025
Department of Oral and Maxillofacial Surgery, Section of Dentistry, Seoul National University Bundang Hospital, 172 Dolma-ro, Bundang-gu, Seongnam-si, Gyeonggi-do, 13620, Republic of Korea.
Background: Traditionally, dental implants require a healing period of 4 to 9 months for osseointegration, with longer recovery times considered when bone grafting is needed. This retrospective study evaluates the clinical efficacy of demineralized dentin matrix (DDM) combined with recombinant human bone morphogenetic protein-2 (rhBMP-2) during dental implant placement to expedite the osseointegration period for early loading.
Methods: Thirty patients (17 male, 13 female; mean age 55.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!