Purpose: To quantify and verify the dosimetric impact of high-dose rate (HDR) source positional uncertainty in brachytherapy, and to introduce a model for three-dimensional (3D) position tracking of the HDR source based on a two-dimensional (2D) measurement. This model has been utilized for the development of a comprehensive source quality assurance (QA) method using radiochromic film (RCF) dosimetry including assessment of different digitization uncertainties.

Methods: An algorithm was developed and verified to generate 2D dose maps of the mHDR-V2 Ir source (Elekta, Veenendaal, Netherlands) based on the AAPM TG-43 formalism. The limits of the dosimetric error associated with source (0.9 mm diameter) positional uncertainty were evaluated and experimentally verified with EBT3 film measurements for 6F (2.0 mm diameter) and 4F (1.3 mm diameter) size catheters at the surface (4F, 6F) and 10 mm further (4F only). To quantify this uncertainty, a source tracking model was developed to incorporate the unique geometric features of all isodose lines (IDLs) within any given 2D dose map away from the source. The tracking model normalized the dose map to its maximum, then quantified the IDLs using blob analysis based on features such as area, perimeter, weighted centroid, elliptic orientation, and circularity. The Pearson correlation coefficients (PCCs) between these features and source coordinates (x, y, z, θ , θ ) were calculated. To experimentally verify the accuracy of the tracking model, EBT3 film pieces were positioned within a Solid Water® (SW) phantom above and below the source and they were exposed simultaneously.

Results: The maximum measured dosimetric variations on the 6F and 4F catheter surfaces were 39.8% and 36.1%, respectively. At 10 mm further, the variation reduced to 2.6% for the 4F catheter which is in agreement with the calculations. The source center (x, y) was strongly correlated with the low IDL-weighted centroid (PCC = 0.99), while the distance to source (z) was correlated with the IDL areas (PCC = 0.96) and perimeters (PCC = 0.99). The source orientation θ was correlated with the difference between high and low IDL-weighted centroids (PCC = 0.98), while θ was correlated with the elliptic orientation of the 60-90% IDLs (PCC = 0.97) for a maximum distance of z = 5 mm. Beyond 5 mm, IDL circularity was significant, therefore limiting the determination of θ (PCC ≤ 0.48). The measured positional errors from the film sets above and below the source indicated a source position at the bottom of the catheter (-0.24 ± 0.07 mm).

Conclusions: Isodose line features of a 2D dose map away from the HDR source can reveal its spatial coordinates. RCF was shown to be a suitable dosimeter for source tracking and dosimetry. This technique offers a novel source QA method and has the potential to be used for QA of commercial and customized applicators.

Download full-text PDF

Source
http://dx.doi.org/10.1002/mp.14540DOI Listing

Publication Analysis

Top Keywords

source
18
hdr source
16
source tracking
12
tracking model
12
dose map
12
tracking hdr
8
quality assurance
8
radiochromic film
8
positional uncertainty
8
ebt3 film
8

Similar Publications

Official development agencies are increasingly supporting civil society lobby and advocacy (L&A) to address poverty and human rights. However, there are challenges in evaluating L&A. As programme objectives are often to change policies or practices in a single institution like a Government Ministry, L&A programmes are often not amenable to large-n impact evaluation methods.

View Article and Find Full Text PDF

Beyond the Herald Patch: Exploring the Complex Landscape of Pityriasis Rosea.

Am J Clin Dermatol

January 2025

Dr. Phillip Frost Department of Dermatology and Cutaneous Surgery, University of Miami Miller School of Medicine, Miami, FL, 33136, USA.

Pityriasis rosea (PR) is a prevalent dermatological condition characterized by a distinctive herald patch, followed by secondary eruptions, often forming a "Christmas tree" pattern on the trunk. Despite its recognizable clinical presentation, the etiology of PR remains uncertain, with hypotheses pointing to both infectious and noninfectious origins. Human herpesviruses (HHV) 6 and 7 have been implicated, with evidence suggesting viral reactivation as a potential trigger.

View Article and Find Full Text PDF

Background: Forecasting future public pharmaceutical expenditure is a challenge for healthcare payers, particularly owing to the unpredictability of new market introductions and their economic impact. No best-practice forecasting methods have been established so far. The literature distinguishes between the top-down approach, based on historical trends, and the bottom-up approach, using a combination of historical and horizon scanning data.

View Article and Find Full Text PDF

Deciphering the colostral-immunity transfer: from mammary gland to neonates small intestine.

Vet Res Commun

January 2025

Laboratory of Agricultural Animal Genetics, Breeding and Reproduction of Ministry of Education, Huazhong Agricultural University, Wuhan, China.

Colostrum, the initial mammary secretion produced by various mammals following birth, is a conduit for maternal immunity transfer in diverse mammalian species. Concurrently, many cellular processes are occurring in the neonatal small intestine to prepare it to receive molecular signals from a superfood essential for the neonate's health and development. During the prepartum colostrum secretion, the newborn intestine undergoes transient alterations in the intestinal barrier, primarily regulating immunoglobulin absorption.

View Article and Find Full Text PDF

Nitrogen doped Carbon Quantum Dots (NCQDs) have been synthesized using most economical and easiest hydrothermal process. Here, N-phenyl orthophenylenediamine and citric acid were utilised as a source of nitrogen and carbon for the preparation of NCQDs. The synthesized NCQDs were characterized using experimental techniques like UV - Vis absorption, FT-IR, transmission electron microscopy (TEM), X-ray Diffraction (XRD), EDX, dynamic light scattering (DLS), fluorimeter and time resolved fluorescence spectroscopy.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!