Background: As serologic assays for SARS-CoV-2 become more widely utilized, it is important to understand their performance characteristics and correlation with neutralizing antibodies. We evaluated 3 commonly used SARS-CoV-2 IgG assays (Abbott, DiaSorin, and EUROIMMUN) for clinical sensitivity, specificity, and correlation with neutralizing antibodies, and then compared antibody kinetics during the acute phase of infection.
Methods: Three panels of samples were tested on every assay. Sensitivity was assessed using a panel of 35 specimens serially collected from 7 patients with RT-PCR-confirmed COVID-19. Specificity was determined using 100 sera samples collected in 2018 from healthy individuals prior to the outbreak. Analytical specificity was determined using a panel of 37 samples from individuals with respiratory illnesses other than COVID-19.
Results: Clinical sensitivity was 91.43% (95% CI 76.94-98.20%) for Abbott, and 88.57% (95% CI 73.26-96.80%) for both DiaSorin and EUROIMMUN. Clinical specificity was 99.00% (95% CI 94.55-99.97%) for Abbott and DiaSorin and 94.00% (95% CI 87.40-97.77%) for EUROIMMUN. The IgG assays demonstrated good qualitative agreement (minimum of 94%) and good correlation between the quantitative result for each combination of assays (r2 ≥ 0.90). The neutralizing antibody response did not necessarily follow the same temporal kinetics as the IgG response and did not necessarily correlate with IgG values.
Conclusion: The 3 IgG antibody assays demonstrated comparable performance characteristics. Importantly, a qualitative positive IgG result obtained with any of the assays was associated with the presence of neutralizing antibodies; however, neutralizing antibody concentrations did not correlate well with signal to cutoff ratios.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7665531 | PMC |
http://dx.doi.org/10.1093/jalm/jfaa188 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!