Requirements for a reliable use of liquid membrane microelectrodes are discussed in terms of stability, response time, and lifetime on the basis of membrane technological considerations. The selectivity of H+, Li+, Na+, K+, Mg2+, Ca2+, and Cl- microelectrodes is critically evaluated using the Nikolskii-Eisenman formalism. Recent progress in the design of new ionophores is presented. A novel neutral carrier-based Ca2+-selective microelectrode with a detection limit of about 5 X 10(-10) M Ca2+ at a background of 125 mM K+ has been realized. An neutral carrier-based microelectrode for H+ with extended pH range of the sample solution is now available. Promising developments in the field of Li+-, Mg2+-, and Cl--selective ionophores are discussed.

Download full-text PDF

Source
http://dx.doi.org/10.1139/y87-141DOI Listing

Publication Analysis

Top Keywords

design ionophores
8
neutral carrier-based
8
ionophores ion-selective
4
ion-selective microsensors
4
microsensors requirements
4
requirements reliable
4
reliable liquid
4
liquid membrane
4
membrane microelectrodes
4
microelectrodes discussed
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!