Delayed entry of patient blood culture samples into a microbial detection system is unavoidable at times, due to off-shift staffing or transporting samples to centralized laboratories. Pre-incubation time and temperature of blood culture bottles are the most critical factors impacting recovery and detection of microorganisms. A total of 1377 BACT/ALERT® (BTA) Fastidious Antimicrobial Neutralization (FAN® PLUS) bottles (FA PLUS, FN PLUS, and PF PLUS) were tested after delayed entry times of 24 and 36 h at 20-25 °C (room temperature, RT) prior to loading into the BACT/ALERT® VIRTUO® microbial detection system (VIRTUO). Clinically relevant organisms were inoculated into bottles with 5-84 colony forming units (CFU) per bottle, and human blood (0 to 10 mL), and then loaded into the VIRTUO. When bottles were loaded without delay, a mean time to detection (TTD) of 9.6 h was observed. For delayed bottles, the TTD reported by the VIRTUO was added to the 24-h and 36-h delay times and resulted in average time to results of 32.5 h and 42.5 h, respectively. The FAN PLUS bottles in conjunction with the VIRTUO produced acceptable results when delays up to 24 h at 20-25 °C occur in loading.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7979663PMC
http://dx.doi.org/10.1007/s10096-020-04042-zDOI Listing

Publication Analysis

Top Keywords

delayed entry
12
blood culture
12
fan bottles
8
microbial detection
8
detection system
8
bottles
7
virtuo
5
delayed
4
entry performance
4
performance bact/alert
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!