Background: Factory-calibrated intermittently-scanned Continuous Glucose Monitoring (isCGM) device FreeStyle Libre (FSL) has recently received improvements in its glucose tracking algorithm and calibration procedures, which are claimed to have improved its accuracy.
Objective: To compare the accuracy of two generations of 14-days FSL devices (A in 2016, B in 2019) to self-monitored blood glucose measurements (SMBG) in children with type 1 diabetes in real-life conditions during a summer camp.
Materials And Methods: Two largely independent groups of youth with type 1 diabetes took part in summer camps. In 2016 they used FSL-A, in 2019 FSL-B. On scheduled days, participants performed supervised 8-point glucose profiles with FSL and SMBG. The accuracy vs SMBG was assessed with mean absolute relative difference (MARD) and clinical surveillance error grid (SEG).
Results: We collected 1655 FSL-SMBG measurement pairs from 78 FSL-A patients (age 13 ± 2.3 years old; HbA1c: 7.6 ± 0.8%) and 1796 from 58 in FSL-B group (age 13.8 ± 2.3 years old, HbA1c: 7.5 ± 1.1%)-in total 3451 measurements. FSL-B displayed lower MARD than FSL-A (11.3 ± 3.1% vs 13.7 ± 4.6%, P = .0003), lower SD of errors (20.2 ± 6.7 mg/dL vs 24.1 ± 9.6 mg/dL, P = .0090) but similar bias (-7.6 ± 11.8 mg/dL vs -6.5 ± 8 mg/dL, P = .5240). Both FSL-A and FSL-B showed significantly higher MARD when glycaemia was decreasing >2 mg/dL/min (FSL-A:22.3 ± 18.5%; FSL-B:17.9 ± 15.8%, P < .0001) compared with stable conditions (FSL-A: 11.4 ± 10.4%, FSL-B:10.1 ± 9.1%) and when the system could not define the glycaemic trend (FSL-A:16.5 ± 16.3%; FSL-B:15.2 ± 14.9%, P < .0001). Both generations demonstrated high percentage of A-class and B-class results in SEG (FSL-A: 96.4%, FSL-B: 97.6%) with a significant shift from B (decrease by 3.7%) to A category (increase by 3.9%) between generations (FSL-A: 16/80.4%; FSL-B:12.3/85.3%, P = .0012).
Conclusion: FSL-B demonstrated higher accuracy when compared to FSL-A However, when glycemia is decreasing or its trend is uncertain, the verification with a glucose meter is still advisable.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1111/pedi.13135 | DOI Listing |
Syst Rev
January 2025
Centre for Clinical Intervention Research, Copenhagen Trial Unit, Capital Region of Denmark, Copenhagen, Denmark.
Background: Type 1 diabetes is a serious, chronic disorder with an increasing incidence among children and adolescents. Glycemic control in individuals with type 1 diabetes is better managed through a basal-bolus regimen with either regular human or rapid-acting insulin analogues administered as a bolus at mealtimes. Rapid-acting insulin analogues have been hypothesized to cause optimal glycemic control and less risk of hypoglycemic episodes compared to regular human insulins.
View Article and Find Full Text PDFDiabetes Obes Metab
January 2025
Department of Endocrine and Metabolic Diseases, Shanghai Institute of Endocrine and Metabolic Diseases, Ruijin Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai, China.
Aim: To explore the holistic impact of socioeconomic and mental health inequalities on the global burden of type 2 diabetes.
Materials And Methods: This cross-sectional study used data on the incidence, disability-adjusted life years (DALYs), and mortality of type 2 diabetes as well as DALYs attributable to risk factors during 1990-2021 from the Global Burden of Disease Study 2021. Average annual percent change (AAPC) was applied to assess the temporal trends from 1990 to 2021.
Diabetes Obes Metab
January 2025
Eli Lilly and Company, Indianapolis, Indiana, USA.
Aims: To explore the relationship between weight loss and insulin sensitivity in response to tirzepatide or semaglutide.
Materials And Methods: We conducted a post hoc exploratory analysis of a 28-week, double-blind, randomized trial in people with type 2 diabetes treated with metformin, randomized to tirzepatide 15 mg, semaglutide 1 mg or placebo. We evaluated the relationship between change in body weight and change in insulin sensitivity determined from hyperinsulinemic euglycemic clamp (M value), or from mixed-meal tolerance testing (Matsuda index).
Diabetes Obes Metab
January 2025
Department of Clinical and Biomedical Sciences, University of Exeter Medical School, Exeter, UK.
Aims: To assess outcomes of oral anti-hyperglycaemic therapies in people with diabetes secondary to a pancreatic condition (type 3c), where specific treatment guidance is limited.
Materials And Methods: Using hospital-linked UK primary care records (Clinical Practice Research Datalink; 2004-2020), we identified 7084 people with a pancreatic condition (acute pancreatitis, chronic pancreatitis, pancreatic cancer and haemochromatosis) preceding diabetes diagnosis (type 3c cohort), initiating oral glucose-lowering therapy (metformin, sulphonylureas, SGLT2-inhibitors, DPP4-inhibitors or thiazolidinediones), and without concurrent insulin treatment. We stratified by pancreatic exocrine insufficiency [PEI] (n = 5917 without PEI, 1167 with PEI) and matched to 97 227 type 2 diabetes (T2D) controls.
J Ovarian Res
January 2025
Departments of Endocrinology, Sheri Kashmir Institute of Medical Sciences, Srinagar, J&K, India.
Background: A significant overlap in the pathophysiological features of polycystic ovary syndrome (PCOS) and type 2 diabetes mellitus (T2DM) has been reported; and insulin resistance is considered a central driver in both. The expression and hepatic clearance of insulin and subsequent glucose homeostasis are mediated by TCF7L2 via Wnt signaling. Studies have persistently associated TCF7L2 genetic variations with T2DM, however, its results on PCOS are sparse and inconsistent.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!