  • The study evaluated the image quality of low iodine concentration dual-energy CT (DECT) with a deep learning noise reduction technique for pediatric abdominal imaging, compared to standard single-energy polychromatic CT (SECT).
  • The results indicated that while DECT had lower attenuation and higher noise than SECT, the combination of deep learning and noise reduction methods maintained overall image quality, especially in liver imaging.
  • DECT also demonstrated significant reductions in radiation dose (19.6%) and total iodine administration (14.3%) compared to SECT, making it a potentially safer option for pediatric patients.

Article Abstract

Objectives: To evaluate the image quality of low iodine concentration, dual-energy CT (DECT) combined with a deep learning-based noise reduction technique for pediatric abdominal CT, compared with standard iodine concentration single-energy polychromatic CT (SECT).

Methods: From December 2016 to May 2017, DECT with 300 mg•I/mL contrast medium was performed in 29 pediatric patients (17 boys, 12 girls; age, 2-19 years). The DECT images were reconstructed using a noise-optimized virtual monoenergetic reconstruction image (VMI) with and without a deep learning method. SECT images with 350 mg•I/mL contrast medium, performed within the last 3 months before the DECT, served as reference images. The quantitative and qualitative parameters were compared using paired t tests and Wilcoxon signed-rank tests, and the differences in radiation dose and total iodine administration were assessed.

Results: The linearly blended DECT showed lower attenuation and higher noise than SECT. The 60-keV VMI showed an increase in attenuation and higher noise than SECT. The combined 60-keV VMI plus deep learning images showed low noise, no difference in contrast-to-noise ratios, and overall image quality or diagnostic image quality, but showed a higher signal-to-noise ratio in the liver and lower enhancement of lesions than SECT. The overall image and diagnostic quality of lesions were maintained on the combined noise reduction approach. The CT dose index volume and total iodine administration in DECT were respectively 19.6% and 14.3% lower than those in SECT.

Conclusion: Low iodine concentration DECT, combined with deep learning in pediatric abdominal CT, can maintain image quality while reducing the radiation dose and iodine load, compared with standard SECT.

Key Points: • An image noise reduction approach combining deep learning and noise-optimized virtual monoenergetic image reconstruction can maintain image quality while reducing radiation dose and iodine load. • The 60-keV virtual monoenergetic image reconstruction plus deep learning images showed low noise, no difference in contrast-to-noise ratio, and overall image quality, but showed a higher signal-to-noise ratio in the liver and a lower enhancement of lesion than single-energy polychromatic CT. • This combination could offer a 19.6% reduction in radiation dose and a 14.3% reduction in iodine load, in comparison with a control group that underwent single-energy polychromatic CT with the standard protocol.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00330-020-07349-9DOI Listing

Publication Analysis

Top Keywords

deep learning
24
image quality
24
noise reduction
16
radiation dose
16
reduction approach
12
pediatric abdominal
12
iodine concentration
12
single-energy polychromatic
12
virtual monoenergetic
12
iodine load
12

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!