Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Introduction: Ceramic-on-ceramic (COC) total hip arthroplasty (THA) was developed to reduce wear debris, and consequently the occurrence of osteolysis and aseptic loosening. The purpose of the present study was to determine the incidence and natural history of noise including squeaking after COC THA, and to clarify whether clinical outcomes and any demographic and implant-related factors are associated with noise from these ceramic bearings.
Methods: This was a retrospective observational study of 137 primary COC THAs between 2004 and 2009 at our institute. The Mean follow-up was for 10.5 years. All patients were followed up with an office visit to obtain information regarding noise, pain, and function. The correlations between the noise, and patient demographic, radiographic, and implant-related factors were evaluated.
Results: Noise developed in 30 hips (21.9%), specifically squeaking in 2 (1.5%) of the 137 hips during the follow-up period. Noise including squeaking was not associated with pain or functional outcomes. There was significant correlation between noise and cup anteversion, in that cup anteversion of patients with noise was smaller than that in those without noise. No significant correlation was found between noise and any other patient demographic including age, sex, height, body weight, or body mass index. Kaplan-Meier survivorship analysis with revision for any reason and an aseptic loosening of either component as the end point, revealed a cumulative survival rate at 10 years of 98.4% and 99.1% respectively.
Conclusion: The long-term clinical and radiographic outcome of cementless COC THA is excellent, but we should be vigilant for noise from COC bearings.