Background: Bone stock preservation in total hip replacement is essential to allow for later revisions in an increasingly younger and fitter index patient population. While contemporary modular press-fit acetabular cups lead to rigid fixation with equatorial stress and central osteolysis, more elastic fixation may cause different peri-acetabular bone remodelling. The purpose of this study was to investigate changes in peri-acetabular bone mineral density (BMD) in uncemented elastic fixation with monoblock press-fit cups.
Methods: This prospective observational study included 45 patients with monoblock cups. We evaluated peri-acetabular BMD using dual-energy X-ray absorptiometry and reported functional outcomes and complications.
Results: At a mean follow-up of 24.2 ± 2.2 months, we found that BMD stabilised in DeLee and Charnley zones I and III and recovered to baseline value in zone II. The mean Harris Hip Scores improved significantly from 56.9 ± 20.0 to 97.2 ± 4.0 (p < 0001). Other than one peri-operative dislocation, we saw no post-operative complications.
Conclusions: We found favourable adaptive bone changes with BMD stabilisation in the equatorial zones and recovery to pre-operative values in the central zone. Additionally, excellent clinical outcomes and few prosthesis-related complications strengthened the favourable results of monoblock acetabular cups.
Trial Registration: Registration number DRKS00017076.