Purpose: Radiation-absorbing pads are an additional possibility to reduce scattered radiation at its source. The goal of this study is to investigate the efficacy of a new reusable radiation-absorbing pad at its origin in an experimental setup.
Material And Methods: All measurements were carried out using a clinical angiography system with a standardized fluoroscopy protocol, different C-arm angulations and an anthropomorphic torso phantom as a scattering body. An ionization chamber was used to measure the radiation exposure at five different heights of a simulated operator during a simulated transfemoral angiography intervention. Measurements were carried out with and without radiation-absorbing pads with lead equivalents of 0.25 and 0.5 mm placed onto the scattering body. For all measurements a mobile acrylic shield and an under-table lead curtain was used.
Results: At all operator heights from 100 to 165 cm a significant radiation dose reduction of up to 80.6 % (p < 0.01) using the radiation-absorbing pad was measured, when compared to no radiation-absorbing pad. At the height of 165 cm the radiation-absorbing pad with a lead equivalence of 0.5 mm showed a significant radiation dose reduction (51.4 %, p < 0.01) in comparison to a lead equivalence of 0.25 mm.
Conclusion: The addition of a radiation-absorbing pad to the standard protection means results in a significant dose reduction for the operator, particularly for upper body parts.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.ejrad.2020.109245 | DOI Listing |
Int J Radiat Biol
October 2023
Department of Physics, Faculty of Science, Atatürk University, Erzurum, Turkey.
Purpose: Determination of the protective property of melanin, an organic polymer class consisting of phenolic and/or indolic compounds isolated from bacteria and fungi, against fast neutron radiation. To show that these melanin samples, which also have antioxidant and metal chelating properties, can be used as an active ingredient for a drug to be developed against neutrons used in nuclear research and medicine.
Materials And Methods: Bacterial and fungal media were prepared, and melanin pigments were produced and isolated.
Clin Res Cardiol
May 2023
Department of Cardiology, University Hospital Zurich, University Heart Center, Raemistrasse 100, 8091, Zurich, Switzerland.
Background: As numbers and complexity of percutaneous coronary interventions are constantly increasing, optimal radiation protection is required to ensure operator safety. Suspended radiation protection systems (SRPS) and protective scatter-radiation absorbing drapes (PAD) are novel methods to mitigate fluoroscopic scattered radiation exposure. The aim of the study was to investigate the effectiveness regarding radiation protection of a SRPS and a PAD in comparison with conventional protection.
View Article and Find Full Text PDFAnn Vasc Surg
August 2022
Department of Vascular Surgery, University Medical Center Utrecht, Utrecht, the Netherlands.
Background: A chronic exposure to low dose radiation, as encountered in endovascular procedures, may impact the health of surgeons and radiologists over a timespan of several months to a lifetime. This study evaluates the feasibility and efficacy of a radiation absorbing sterile drape (RADPAD) to reduce operator exposure during the endovascular treatment of obstructive peripheral artery disease (PAD).
Methods: Between February 2016 and September 2017, patients with PAD who received percutaneous transluminal angioplasty, stent placement, remote endarterectomy, or a combination thereof were included in this nonrandomized study.
Catheter Cardiovasc Interv
March 2022
Division of Cardiology, Frederik Meijer Heart and Vascular Institute, Spectrum Health, Grand Rapids, Michigan, USA.
Objective: This study was performed to evaluate physician radiation doses with the use of a suspended lead suit.
Background: Interventional cardiologists face substantial occupational risks from chronic radiation exposure and wearing heavy lead aprons.
Methods: Head-level physician radiation doses, collected using real-time dosimeters during consecutive coronary angiography procedures, were compared with the use of a suspended lead suit versus conventional lead aprons.
Eur J Radiol
November 2020
Philipps-University of Marburg, Clinic of Diagnostic and Interventional Radiology, University Hospital Marburg, Germany.