A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 176

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML

File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 316
Function: require_once

Analytical performance evaluation of a commercial next generation sequencing liquid biopsy platform using plasma ctDNA, reference standards, and synthetic serial dilution samples derived from normal plasma. | LitMetric

Background: Circulating tumor (ct) DNA assays performed in clinical laboratories provide tumor biomarker testing support for biopharmaceutical clinical trials. Yet it is neither practical nor economically feasible for many of these clinical laboratories to internally develop their own liquid biopsy assay. Commercially available ctDNA kits are a potential solution for laboratories seeking to incorporate liquid biopsy into their test menus. However, the scarcity of characterized patient samples and cost of purchasing validation reference standards creates a barrier to entry. In the current study, we evaluated the analytical performance of the AVENIO ctDNA liquid biopsy platform (Roche Sequencing Solutions) for use in our clinical laboratory.

Method: Intra-laboratory performance evaluation of AVENIO ctDNA Targeted, Expanded, and Surveillance kits (Research Use Only) was performed according to College of American Pathologists (CAP) guidelines for the validation of targeted next generation sequencing assays using purchased reference standards, de-identified human plasma cell-free (cf) DNA samples, and contrived samples derived from commercially purchased normal and cancer human plasma. All samples were sequenced at read depths relevant to clinical settings using the NextSeq High Output kit (Illumina).

Results: At the clinically relevant read depth, Avenio ctDNA kits demonstrated 100% sensitivity in detecting single nucleotide variants (SNVs) at ≥0.5% allele frequency (AF) and 50% sensitivity in detecting SNVs at 0.1% AF using 20-40 ng sample input amount. The assay integrated seamlessly into our laboratory's NGS workflow with input DNA mass, target allele frequency (TAF), multiplexing, and number of reads optimized to support a high-throughput assay appropriate for biopharmaceutical trials.

Conclusions: Our study demonstrates that AVENIO ctDNA liquid biopsy platform provides a viable alternative for efficient incorporation of liquid biopsy assays into the clinical laboratory for detecting somatic alterations as low as 0.5%. Accurate detection of variants lower than 0.5% could potentially be achieved by deeper sequencing when clinically indicated and economically feasible.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7528227PMC
http://dx.doi.org/10.1186/s12885-020-07445-5DOI Listing

Publication Analysis

Top Keywords

liquid biopsy
24
avenio ctdna
16
biopsy platform
12
reference standards
12
analytical performance
8
performance evaluation
8
generation sequencing
8
samples derived
8
clinical laboratories
8
economically feasible
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!