Reclassification of prosthesis-patient mismatch after transcatheter aortic valve replacement using predicted vs. measured indexed effective orifice area.

Eur Heart J Cardiovasc Imaging

Institut Universitaire de Cardiologie et de Pneumologie de Québec- Université Laval/Québec Heart and Lung Institute, Laval University, 2725 Chemin Sainte-Foy, Québec city, Québec G1V-4G5, Canada.

Published: January 2021

Aims: The objective was to compare the incidence and impact on outcomes of measured (PPMM) vs. predicted (PPMP) prosthesis-patient mismatch following transcatheter aortic valve replacement (TAVR).

Methods And Results: All consecutives patients who underwent TAVR between 2007 and 2018 were included. Effective orifice area (EOA) was measured by Doppler-echocardiography using the continuity equation and predicted according to the normal reference for each model and size of valve. PPM was defined using EOA indexed (EOAi) to body surface area as moderate if ≤0.85 cm2/m2 and severe if ≤ 0.65 cm2/m2 (respectively, ≤ 0.70 and ≤ 0.55 cm2/m2 if body mass index ≥ 30 kg/m2). The outcome endpoints were high residual gradient (≥20 mmHg) and the composite of cardiovascular mortality and hospital readmission for heart failure at 1 year. Overall, 1088 patients underwent a TAVR (55% male, age 79.1 ± 8.4 years, and STS score 6.6 ± 4.7%); balloon-expandable device was used in 83%. Incidence of moderate (10% vs. 27%) and severe (1% vs. 17%) PPM was markedly lower when defined by predicted vs. measured EOAi (P < 0.001). Balloon-expandable device implantation (OR: 1.90, P = 0.029) and valve-in-valve procedure (n = 118; OR: 3.21, P < 0.001) were the main factors associated with PPM occurrence. Compared with measured PPM, predicted PPM showed stronger association with high residual gradient. Severe measured or predicted PPM was not associated with clinical outcomes.

Conclusion: The utilization of the predicted EOAi reclassifies the majority of patients with PPM to no PPM following TAVR. Compared with measured PPM, predicted PPM had stronger association with haemodynamic outcomes, while both methods were not associated with clinical outcomes.

Download full-text PDF

Source
http://dx.doi.org/10.1093/ehjci/jeaa235DOI Listing

Publication Analysis

Top Keywords

prosthesis-patient mismatch
8
mismatch transcatheter
8
transcatheter aortic
8
aortic valve
8
valve replacement
8
predicted measured
8
effective orifice
8
orifice area
8
patients underwent
8
underwent tavr
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!