Purpose: To investigate whether a novel iterative cone-beam computed tomography (CBCT) reconstruction algorithm reduces metal artifacts in head and neck patient images.

Method: An anthropomorphic phantom and 35 patients with dental metal prostheses or implants were analyzed. All CBCT images were acquired using a TrueBeam linear accelerator and reconstructed with a Feldkamp-Davis-Kress algorithm-based CBCT (FDK-CBCT) and an iterative CBCT algorithm. The mean Hounsfield unit (HU) and standard deviation values were measured on the tongue near the metal materials and the unaffected region as reference values. The artifact index (AI) was calculated. For objective image analysis, the HU value and AI were compared between FDK-CBCT and iterative CBCT images in phantom and clinical studies. Subjective image analyses of metal artifact scores and soft tissue visualizations were conducted using a five-point scale by two reviewers in the clinical study.

Results: The HU value and AI showed significant artifact reduction for the iterative CBCT than for the FDK-CBCT images (phantom study: 389.8 vs.-10.3 for HU value, 322.9 vs. 96.2 for AI, FDK-CBCT vs. iterative CBCT, respectively; clinical study: 210.3 vs. 69.0 for HU value, 149.6 vs. 70.7 for AI). The subjective scores in the clinical patient study were improved in the iterative CBCT images (metal artifact score: 1.1 vs. 2.9, FDK-CBCT vs. iterative CBCT, respectively; soft tissue visualization: 1.8 vs. 3.6).

Conclusions: The iterative CBCT reconstruction algorithm substantially reduced metal artifacts caused by dental metal prostheses and improved soft tissue visualization compared to FDK-CBCT in phantom and clinical studies.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ejrad.2020.109293DOI Listing

Publication Analysis

Top Keywords

iterative cbct
32
fdk-cbct iterative
16
metal artifact
12
cbct reconstruction
12
reconstruction algorithm
12
phantom clinical
12
cbct images
12
soft tissue
12
cbct
11
iterative
9

Similar Publications

Noise-matched total-likelihood-based bilateral filter: Experimental feasibility in a benchtop photon-counting CBCT system.

Phys Med

January 2025

Department of Robotics and Mechatronics Engineering, Daegu Gyeongbuk Institute of Science and Technology (DGIST), 333 Techno Jungang-daero, Hyeonpung-eup, Dalseong-gun, Daegu, 42988, Republic of Korea. Electronic address:

Purpose: Material decomposition induces substantial noise in basis images and their synthesized computed tomography (CT) images. A likelihood-based bilateral filter was previously developed as a neighborhood filter that effectively reduces noise. However, this method is sensitive to image contrast, and the noise texture needs improvement.

View Article and Find Full Text PDF
Article Synopsis
  • Limited-angle dual-energy cone-beam CT (LA-DECBCT) is a promising method for achieving fast, low-dose imaging, but its clinical use is challenged by difficulties in image reconstruction.
  • A new image reconstruction technique using inter-spectral structural similarity was developed to reduce artifacts, improving the quality of DECBCT images without needing extra data for training.
  • This method shows significant potential for practical clinical applications in LA-DECBCT, enabling accurate imaging without relying on X-ray spectra or paired datasets.
View Article and Find Full Text PDF
Article Synopsis
  • The study aims to improve visualization of arteries during endovascular procedures for peripheral artery disease by using an image registration technique that fuses X-ray and CT angiography images.
  • The method involved aligning digital images based on the positions of the bones and achieved successful registration in most cases, with accurate alignment of less than 1 mm in distance.
  • The results indicate that this technique is clinically viable for guiding interventions, as it allows for early detection of potential complications like guidewire perforations while maintaining a reasonable processing time.
View Article and Find Full Text PDF

Many tasks performed in image-guided procedures can be cast as pose estimation problems, where specific projections are chosen to reach a target in 3D space. In this study, we construct a framework for fluoroscopic pose estimation and compare alternative loss functions and volumetric scene representations. We first develop a differentiable projection (DiffProj) algorithm for the efficient computation of Digitally Reconstructed Radiographs (DRRs) from either Cone-Beam Computerized Tomography (CBCT) or neural scene representations.

View Article and Find Full Text PDF

A comprehensive dual energy method for CBCT metal artifact reduction.

Phys Med Biol

December 2024

Department of Engineering and Applied Physics, University of Science and Technology of China, Hefei, Anhui 230026, People's Republic of China.

A major limitation in cone beam CT (CBCT) application is the presence of metal artifacts when scanning metal-embedded objects or high attenuation materials. This study aims to develop a dual-energy based method for effective metal artifact reduction.The proposed method comprised three steps.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!