ARC Training Centre for Portable Analytical Separation Technologies (ASTech), School of Natural Sciences, University of Tasmania, Private Bag 75, Hobart, Tasmania 7001, Australia; Australian Centre for Research on Separation Science (ACROSS), School of Natural Sciences, University of Tasmania, Private Bag 75, Hobart, Tasmania 7001, Australia; ARC Centre of Excellence for Electromaterials Science (ACES), School of Natural Sciences, University of Tasmania, Private Bag 75, Hobart, Tasmania 7001, Australia. Electronic address:

Published: November 2020

A new miniaturised capillary flow-through deep-UV absorbance detector has been developed using a microscale surface mounted device- type light-emitting diode (LED) (Crystal IS OPTAN 3535-series), emitting at 235 nm and with a half-height band width of 12 nm, and a high-sensitivity Z-shaped flow-cell. Compared with a previously reported TO-39 ball lens LEDs emitting at 235 nm, the new generation LED produced a 20-fold higher optical output and delivered up to 35 times increase in external quantum efficiency (EQE). The Z-cell was based on a reflective rectangular optical path with cross-sectional dimensions of 100 × 100 µm and a physical optical pathlength of 1.2 mm. Inclusion of UV transparent fused-silica ball lenses, between the SMD and the Z-cell, improved light transmission by a factor of 9 and improved the detector signal-to-noise ratio by a factor of 2.2, at the same input current. The detector was housed within an Al-housing fitted with a cooling fan and demonstrated excellent linearity with stray light down to 0.06%, and an effective pathlength of 1.1 mm (92% of nominal pathlength). The resultant detector was fitted successfully into a briefcase-sized portable capillary HPLC system, and practically demonstrated with the detection of a mixture of 13 test compounds at the sub-mg L level in <5 min using gradient elution.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.chroma.2020.461540DOI Listing

Publication Analysis

Top Keywords

absorbance detector
8
surface mounted
8
light-emitting diode
8
emitting 235
8
detector
5
ultraviolet absorbance
4
detector based
4
based high
4
high output
4
output power
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!