Background: Numerous biomechanical and clinical studies comparing different techniques for rotator cuff repair have been reported, yet universal consensus regarding the superior technique has not achieved. A medially-based single-row with triple-loaded suture anchor (also referred to as the Southern California Orthopedic Institute [SCOI] row) and a suture-bridge double-row (SB-DR) with Push-Locks have been shown to result in comparable improvement in treating rotator cuff tear, yet the biomechanical difference is unknown. The purpose of the current study was to determine whether a SCOI row repair had comparable initial biomechanical properties to a SB-DR repair.

Methods: Six matched pairs of fresh-frozen cadaveric shoulders with full-thickness supraspinatus tendon tears we created were included. Two different repairs were performed for each pair (SCOI row and SB-DR methods). Specimens were mounted on a material testing machine to undergo cyclic loading, which was cycled from 10 to 100 N at 1 Hz for 500 cycles. Construct gap formation was recorded at an interval of 50 cycles. Samples were then loaded to failure and modes of failure were recorded. Repeated-measures analysis of variance and pair-t test were used for statistical analyses.

Results: The construct gap formation did not differ between SCOI row and SB-DR repairs (P = 0.056). The last gap displacement was 1.93 ± 0.37 mm for SCOI row repair, and 1.49 ± 0.55 mm for SB-DR repair. The tensile load for 5 mm of elongation and ultimate failure were higher for SCOI row repair compared to SB-DR repair (P = 0.011 and 0.028, respectively). The ultimate failure load was 326.34 ± 11.52 N in the SCOI row group, and 299.82 ± 27.27 N in the SB-DR group. Rotator cuff repair with the SCOI row method failed primarily at the suture- tendon interface, whereas pullout of the lateral row anchors was the primary mechanism of failure for repair with the SB-DR method.

Conclusion: Rotator cuff repair with the SCOI row method has superior biomechanical properties when compared with the SB-DR method. Therefore, SCOI row repair using a medially-based single-row technique with triple-loaded suture anchor is recommended to improve the initial strength in treating full-thickness rotator cuff tears.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7517672PMC
http://dx.doi.org/10.1186/s12891-020-03654-yDOI Listing

Publication Analysis

Top Keywords

scoi row
36
rotator cuff
24
cuff repair
16
row repair
16
triple-loaded suture
12
suture anchor
12
repair
11
row
11
sb-dr
9
scoi
9

Similar Publications

Since the introduction of rotator cuff repair in the beginning of the 20th century, there have been significant advances in the surgical techniques and improvement in patient outcomes. However, controversy remains regarding the best method for tendon repair, including implant choice, ideal suture construct, and the potential benefits of supplemental biologic additives in order to achieve repair with the best opportunity for healing. The "SCOI Row" technique has been refined over three decades with extensive application and consistently good outcomes.

View Article and Find Full Text PDF

Background: Arthroscopic repair is recommended for young patients with full-thickness rotator cuff tears (RCTs), but the healing rates have raised concerns. The Southern California Orthopedic Institute (SCOI) row method has been developed based on greater than 3 decades of experience with excellent clinical outcomes; however, studies with a focus on the younger patient population are limited in number. The current study assessed the short-term clinical outcome and the initial tendon-to-bone healing in a young cohort after repair of a full-thickness RCT using the SCOI row method.

View Article and Find Full Text PDF

Background: Numerous biomechanical and clinical studies comparing different techniques for rotator cuff repair have been reported, yet universal consensus regarding the superior technique has not achieved. A medially-based single-row with triple-loaded suture anchor (also referred to as the Southern California Orthopedic Institute [SCOI] row) and a suture-bridge double-row (SB-DR) with Push-Locks have been shown to result in comparable improvement in treating rotator cuff tear, yet the biomechanical difference is unknown. The purpose of the current study was to determine whether a SCOI row repair had comparable initial biomechanical properties to a SB-DR repair.

View Article and Find Full Text PDF

Background: The purpose of this study was to assess the midterm clinical and ultrasonographic outcomes of a new all arthroscopic supraglenoid origin-preserving tenodesis technique of the long head of the biceps (LHB) brachii tendon in the setting of small to medium-sized rotator cuff repairs.

Materials And Methods: Thirty patients (33 shoulders) meeting inclusion criteria were identified who underwent LHB tenodesis with this technique in the setting of small to medium-sized rotator cuff repair at a mean age at surgery of 65.6 years between 2015 and 2017.

View Article and Find Full Text PDF

Background: The number of anchors and modality of fixation to be used has been a point of debate in the arthroscopic management of rotator cuff repair. Southern California Orthopedic Institute (SCOI) technique has shown better anatomical healing of tendons using single-row anchors. In this study, we evaluated the functional outcomes of arthroscopic rotator cuff repair using the SCOI technique, in Indian population.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!