Purpose: To develop fast multi-slice apparent T (T ) mapping for accurate cerebral blood flow (CBF) quantification with arterial spin labeling (ASL) MRI.

Methods: Fast multi-slice T was measured using a modified inversion recovery echo planar imaging (EPI) sequence with simultaneous application of ASL tagging radiofrequency (RF) and gradient pulses. The fast multi-slice T measurement was compared with the single-slice T imaging approach, repeated per slice. CBF was assessed in healthy adult Wistar rats (N = 5) and rats with acute stroke 24 hours after a transient middle cerebral artery occlusion (N = 5).

Results: The fast multi-slice T measurement was in good agreement with that of a single-slice T imaging approach (Lin's concordance correlation coefficient = 0.92). CBF calculated using T reasonably accounted for the finite labeling RF duration, whereas the routine T -normalized ASL MRI underestimated the CBF, particularly at short labeling durations. In acute stroke rats, the labeling time and the CBF difference (ΔCBF) between the contralateral normal area and the ischemic lesion were significantly correlated when using T -normalized perfusion calculation (R = 0.844, P = .035). In comparison, T -normalized ΔCBF had little labeling time dependence based on the linear regression equation of ΔCBF = -0.0247*τ + 1.579 mL/g/min (R = -0.352, P = .494).

Conclusions: Our study found fast multi-slice T imaging improves the accuracy and reproducibility of CBF measurement.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7718392PMC
http://dx.doi.org/10.1002/mrm.28510DOI Listing

Publication Analysis

Top Keywords

fast multi-slice
24
multi-slice apparent
8
apparent mapping
8
arterial spin
8
spin labeling
8
cerebral blood
8
blood flow
8
multi-slice measurement
8
single-slice imaging
8
imaging approach
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!