An enzyme-linked immunosorbent assay for Treponema pallidum antibody (Syphilis Bio-EnzaBead Kit; Organon Teknika Corp., formerly Litton Bionetics, Kensington, Md.) was compared with the standard fluorescent treponemal antibody absorption test for syphilis (indirect fluorescent-antibody confirmatory test; Zeus Scientific Inc., Raritan, N.J.). Six hundred specimens tested in the rapid plasma Reagin card test (Hynson, Westcott and Dunning, Inc., Baltimore, Md.) and microhemagglutination assay for antibodies to T. pallidum (Ames Division, Miles Laboratories, Inc., Elkhart, Ind.) were used for comparison testing. One hundred and sixty-two specimens were from either persons with syphilis or persons whose history was unknown but who showed reactive serology, whereas 438 were from persons without syphilis. The reactivity of each serum by the Bio-EnzaBead test was determined visually and from optical density readings. Excellent reproducibility of visual readings was obtained; all results were within +/- 1 gradation of the expected readings in all 60 coded sera tested. Overall agreement between the Bio-EnzaBead and fluorescent treponemal antibody absorption test results was 96.3% when read visually and greater than or equal to 95.7% when using optical density readings. Our data indicate that Bio-EnzaBead results read at 405 nm and determined by using the mean antigen optical density reading of the nonreactive control plus 0.025 as the cutoff value provide an overall sensitivity and specificity of 93.0 and 98.6%, respectively, along with the best agreement with the fluorescent treponemal antibody absorption assay. Ease of performance and objectivity also contribute toward the acceptability of this assay as an alternative confirmatory test for syphilis.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC266093PMC
http://dx.doi.org/10.1128/jcm.25.5.808-811.1987DOI Listing

Publication Analysis

Top Keywords

treponemal antibody
16
fluorescent treponemal
12
antibody absorption
12
optical density
12
syphilis bio-enzabead
8
absorption test
8
test syphilis
8
confirmatory test
8
persons syphilis
8
density readings
8

Similar Publications

Bedřich Smetana's illness associated with complete hearing loss has been linked to various diseases in a long-running discussion. Senile dementia, atherosclerosis, progressive paralysis in lues are mentioned, while the ototoxicity of mercury is neglected. Its high concentration in the composer's skeletal remains was demonstrated by the authors of this article immunohistochemically and especially by the neutron activation analysis (NAA) method 23 years ago.

View Article and Find Full Text PDF

Background: The challenge of dealing with isolated reactive treponemal chemiluminescence immunoassay (CIA) results in clinical practice has prompted the development of a more efficient algorithm for distinguishing true infection from false reactivity in isolated CIA sera.

Methods: A prospective cohort study was conducted at Wuhan Tongji Hospital, involving 119,002 individuals screened for syphilis using CIA from January 1, 2015, to January 6, 2017. Samples with reactive CIA results underwent simultaneous testing with the T.

View Article and Find Full Text PDF

Laboratory evaluation of oral fluid for syphilis screening among clinic users from Lima, Peru.

Microbiol Spectr

January 2025

Department of Population and Public Health Sciences, Keck School of Medicine, University of Southern California, Los Angeles, California, USA.

Serological tests for syphilis require equipment unavailable in many health centers; however, point-of-care testing facilitates rapid screening using finger-prick whole blood samples. A further improvement could be oral fluid for ease of use. We evaluated the performance of treponemal antibody point-of-care testing for syphilis screening using oral fluid samples.

View Article and Find Full Text PDF

Update on syphilis diagnostics.

Curr Opin Infect Dis

February 2025

Division of Infectious Diseases, Department of Medicine, University of North Carolina, Chapel Hill, North Carolina, USA.

Purpose Of Review: Syphilis is rising globally, with resultant increases in morbidity and mortality that disproportionately impact underserved populations. Appropriate screening and testing for syphilis with accurate, reliable, affordable, and timely assays is crucial for combating this epidemic. This review provides a brief overview of laboratory methods for syphilis diagnosis, with a focus on recent updates in diagnostics and directions for future research.

View Article and Find Full Text PDF
Article Synopsis
  • Syphilis, caused by a spirochetal pathogen, can lead to serious health complications if untreated, necessitating new strategies to understand its complex pathogenesis.
  • This study engineered a strain of the syphilis bacteria that expresses green fluorescent protein (GFP), allowing researchers to visualize its interactions with host cells and evaluate its infectivity in a rabbit model.
  • The findings emphasize the role of specific antibody responses in targeting the bacteria's outer membrane, highlighting potential targets for protective immunity and paving the way for future research on spirochetal-host interactions.
View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!