A Tensionable Suture-based Cerclage Is an Alternative to Stainless Steel Cerclage Fixation for Stabilization of a Humeral Osteotomy During Shoulder Arthroplasty.

J Am Acad Orthop Surg

From Southern Oregon Orthopedics, Medford, OR (Dr. Denard), the Department of Orthopaedics and Rehabilitation, Oregon Health & Science University, Portland, OR (Dr. Denard), Steadman Philippon Research Institute, Vail, CO (Dr. Nolte and Dr. Millett), BG Trauma Center Ludwigshafen at the University of Heidelberg, Clinic for Trauma and Orthopaedic Surgery, Ludwigshafen, Germany (Dr. Nolte), the The Steadman Clinic, Vail, CO (Dr. Millett), Naples Community Hospital (Dr. Adams and Mr. Rego), and Arthrex, Naples, FL (Dr. Adams, Dr. Liebler, Mr. Rego, and Dr. Higgins).

Published: June 2021

Introduction: Fixation of periprosthetic humeral fractures is most commonly obtained with steel-based wires or cables; however, disadvantages with these constructs are numerous. Suture-based cerclages offer the advantage of easy handling, less radiographic interference, and risk of metallosis, as well as decreased risk of cutting into the soft humeral bone. Therefore, the purpose of this study was to compare a suture-based cerclage to a stainless steel wire cerclage (SSWC) for stabilization of the humerus during shoulder arthroplasty.

Methods: In part I of the study, SSWC fixation was compared with single-looped tape cerclage and a double-looped tape cerclage (DLTC) fixation. In part II, a subsidence test was performed on 12 cadaveric humeri. After an osteotomy, the humeri were secured with either a SSWC or DLTC. Subsequently, a metal wedge was introduced into the humerus to simulate the stem of a shoulder arthroplasty.

Results: In part I, load to 2-mm displacement was significantly higher for the DLTC construct compared with the SSWC construct (2,401 ± 483 N versus 750 ± 33 N; P < 0.0001). Load to failure was 935 ± 143 N with the SSWC, 1,737 ± 113 N with the single-looped tape cerclage, and 4,360 ± 463 N with the DLTC constructs, and all differences were statistically significant (P < 0.05). In part II, load at 20-mm subsidence was higher for the DLTC (320 ± 274 N) compared with the SSWC (247 ± 137 N), but no significant difference was observed (P > 0.05). However, gap displacement at 20 mm subsidence was significantly lower with the DLTC construct (0.33 ± 0.31 mm versus 0.77 ± 0.23 mm; P = 0.009). Load to failure was higher with the DLTC construct compared with the SSWC construct (4,447 ± 2,325 N versus 1,880 ± 1,089 N; P = 0.032), but the final gap displacement did not differ significantly (DLTC 5.23 ± 6.63 mm versus SSWC 6.03 ± 8.82 mm; P > 0.05).

Discussion: A DLTC has higher load to failure and trends toward lower gap displacement compared with a SSWC. The DLTC construct may therefore be a viable alternative for fixation of periprosthetic fractures or osteotomies of the humeral shaft during shoulder arthroplasty.

Download full-text PDF

Source
http://dx.doi.org/10.5435/JAAOS-D-20-00047DOI Listing

Publication Analysis

Top Keywords

dltc construct
16
compared sswc
16
tape cerclage
12
higher dltc
12
load failure
12
gap displacement
12
dltc
10
sswc
9
suture-based cerclage
8
stainless steel
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!