Measurement of Glutathione as a Tool for Oxidative Stress Studies by High Performance Liquid Chromatography.

Molecules

Hull York Medical School & Department of Renal Medicine, Hull University Teaching Hospitals Trust, Anlaby Road, Kingston upon, Hull HU3 2JZ, UK.

Published: September 2020

Background: Maintenance of the ratio of glutathione in the reduced (GSH) and oxidised (GSSG) state in cells is important in redox control, signal transduction and gene regulation, factors that are altered in many diseases. The accurate and reliable determination of GSH and GSSG simultaneously is a useful tool for oxidative stress determination. Measurement is limited primarily to the underestimation of GSH and overestimation GSSG as a result of auto-oxidation of GSH. The aim of this study was to overcome this limitation and develop, optimise and validate a reverse-phase high performance liquid chromatographic (HPLC) assay of GSH and GSSG for the determination of oxidant status in cardiac and chronic kidney diseases.

Methods: Fluorescence detection of the derivative, glutathione-O-pthaldialdehyde (OPA) adduct was used. The assay was validated by measuring the stability of glutathione and glutathione-OPA adduct under conditions that could affect the reproducibility including reaction time and temperature. Linearity, concentration range, limit of detection (LOD), limit of quantification (LOQ), recovery and extraction efficiency and selectivity of the method were assessed.

Results: There was excellent linearity for GSH (r = 0.998) and GSSG (r = 0.996) over concentration ranges of 0.1 µM-4 mM and 0.2 µM-0.4 mM respectively. The extraction of GSH from tissues was consistent and precise. The limit of detection for GSH and GSSG were 0.34 µM and 0.26 µM respectively whilst their limits of quantification were 1.14 µM and 0.88 µM respectively.

Conclusion: These data validate a method for the simultaneous measurement of GSH and GSSG in samples extracted from biological tissues and offer a simple determination of redox status in clinical samples.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7571047PMC
http://dx.doi.org/10.3390/molecules25184196DOI Listing

Publication Analysis

Top Keywords

gsh gssg
16
gsh
9
tool oxidative
8
oxidative stress
8
high performance
8
performance liquid
8
limit detection
8
gssg
7
measurement glutathione
4
glutathione tool
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!