Clinical evaluation of a new monofocal IOL with enhanced intermediate function in patients with cataract.

J Cataract Refract Surg

From the International Vision Correction Research Centre (IVCRC), University-Eye-Clinic (Auffarth), Heidelberg, Augenklinik Ahaus (Gerl), Ahaus, Germany; Johnson & Johnson Vision (Tsai, Janakiraman, Jackson), Santa Ana, California, USA; AMO (Alarcon), Groningen, the Netherlands; University Eye Hospital (Dick), Bochum, Germany.

Published: February 2021

Purpose: To evaluate the effectiveness and safety of 2 enhanced monofocal intraocular lenses (IOLs). The TECNIS Eyhance IOL (Model ICB00) was compared with a standard monofocal IOL (TECNIS Monofocal, Model ZCB00).

Setting: European multicenter study.

Design: Prospective, bilateral, randomized, comparative/evaluator-masked, controlled study.

Methods: Adult subjects scheduled to undergo bilateral, primary phacoemulsification cataract extraction and posterior IOL implantation were randomized to receive the enhanced monofocal ICB00 IOL or the monofocal ZCB00 IOL in both eyes. Monocular endpoints at 6 months included distance-corrected intermediate visual acuity (DCIVA), photopic corrected distance visual acuity, and uncorrected intermediate visual acuity (UIVA). Binocular visual acuities, monocular corrected distance contrast sensitivity (first eyes), patient-reported outcomes, and safety were assessed at 6 months.

Results: Overall, 139 patients were bilaterally implanted with the enhanced monofocal IOL (n = 67) or standard monofocal IOL (n = 72) and available for the 6-month visit. The enhanced monofocal IOL significantly improved mean monocular and binocular DCIVA and UIVA by at least 1-line logarithm of the minimum angle of resolution vs the standard monofocal IOL (all P ≤ .0001). Distance vision for the enhanced monofocal IOL was 20/20 or better and comparable with that of the standard monofocal lens at 6 months. Contrast sensitivity, photic phenomena outcomes, and rates of adverse events were similar between the 2 groups.

Conclusions: In patients undergoing cataract surgery, TECNIS Eyhance IOL Model ICB00 provided enhanced intermediate vision and similar distance performance and photic phenomena compared with a standard monofocal IOL, along with improved functional performance in daily life.

Download full-text PDF

Source
http://dx.doi.org/10.1097/j.jcrs.0000000000000399DOI Listing

Publication Analysis

Top Keywords

monofocal iol
32
enhanced monofocal
20
standard monofocal
20
monofocal
13
iol
13
visual acuity
12
enhanced intermediate
8
tecnis eyhance
8
eyhance iol
8
iol model
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!