Dimensional variability of orthodontic slots and archwires: an analysis of torque expression and clinical implications.

Prog Orthod

Department of Biotechnological and Applied Clinical Sciences, University of L'Aquila, Viale S. Salvatore, Edificio Delta 6, 67100, L'Aquila, Italy.

Published: September 2020

AI Article Synopsis

  • The study investigates how different dimensions of orthodontic brackets and archwires affect torsional play, a factor that impacts clinical outcomes.
  • Measurements revealed that bracket slots are typically larger than intended, while archwire sizes can vary, affecting the fit and resulting in torsional play that can hinder effective torque application.
  • The findings emphasize the need for better manufacturer specifications on the edge bevel of archwires, as variability in production can lead to significant differences in clinical performance.

Article Abstract

Background: The loss of third-order information in pre-adjusted brackets due to torsional play is a problem in clinical orthodontics. The aim of this study was to evaluate the impact of slot height, archwire height, width and edge bevel's radius on the torsional play for three brackets/archwire systems.

Methods: Ninety brackets with a 0.022 × 0.028 in. slot with McLaughlin-Bennett-Trevisi prescription from three different manufacturers were selected, and the slot's height and depth were measured using a profile projector. Sixty stainless-steel rectangular archwires from three different manufacturers were sectioned and observed with a SEM to measure their height, width, and radius of edge bevel. The recorded data were used to calculate the theoretical torsional play between different slot-archwire combinations. One-way ANOVA was used to compare the measurements within different bracket types and among different manufacturers.

Results: Slot height was usually oversized. Archwire's height was usually undersized, but oversized wires were also observed. The radius edge bevel was the most variable parameter. A certain degree of torsional play is always present that differs from one bracket type to another of the same producer and that can even be doubled from one manufacturer to another.

Conclusions: Due to production tolerance, differences between the nominal values and the real dimensions of any components of a slot/archwire system are common. This results in a torsional play that limits torque expression. The archwire's edge bevel plays an important role in torque expression, and clearer information should be provided by the manufacturers regarding this aspect.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7487281PMC
http://dx.doi.org/10.1186/s40510-020-00333-5DOI Listing

Publication Analysis

Top Keywords

torsional play
20
torque expression
12
edge bevel
12
slot height
8
height width
8
three manufacturers
8
radius edge
8
height
6
torsional
5
play
5

Similar Publications

The structural plasticity of proteins at the molecular level is largely dictated by backbone torsion angles, which play a critical role in ligand recognition and binding. To establish the anion-induced cooperative arrangement of the main-chain (mc) torsion, herein, we analyzed a set of naturally occurring CαNN motifs as "static models" for their anion-binding competence through docking and molecular dynamics simulations and decoded its torsion angle influenced mc-driven anion recognition potential. By comparing a pool of 20 distinct sets of CαNN motif with identical sequences in their "anion bound/present, aP" and "anion free/absent, aA" versions, we could discern that there exists a positive correlation between the "difference of anion residence time (ΔR)" and "difference among the main-chain torsion angle" of the aP and aA population.

View Article and Find Full Text PDF

Proximal tibial osteotomies for treating grade IV medial patellar luxation in two feline stifles.

Top Companion Anim Med

January 2025

OrtopediaVet Veterinary Clinic, Dr. Eloy Curuci and team. Volta Redonda Street, 670 - Campo Belo, São Paulo, State of São Paulo 04608-011, Brazil.

Objectives: There is a lack of literature regarding the surgical management of high-grade patellar luxation in cats. Among the available options, corrective osteotomies are suitable for correcting severe bone deformities. Therefore, this study aimed to report on the surgical management of grade IV medial patellar luxation (MPL) through tibial corrective osteotomies in two cats.

View Article and Find Full Text PDF

Four new and one previously reported silver 4,4'-vinylenedipyridine (Vpe) coordination polymers were tested as anion exchange materials to assess their potential for pollutant sequestration and compared to analogous silver 4,4'-bipyridine (bipy) coordination polymers. The materials were synthesized using nitrate, tetrafluoroborate, perchlorate, perrhenate, or chromate as the anion to produce cationic coordination polymers with solubilities ranging from 0.0137(7) to 0.

View Article and Find Full Text PDF

The testicular artery is the vessel that provides oxygen-rich blood to the testes. It is also involved in many more processes that play important roles in spermatogenesis such as thermoregulation. As the vessel has a number of anatomical variants, it may be often responsible for difficulties and complications during abdominopelvic surgical procedures.

View Article and Find Full Text PDF

A one-step fabrication of soft-magnetic high entropy alloy fiber with excellent strength and flexibility.

Nat Commun

December 2024

CAS Key Laboratory of Magnetic Materials and Devices, Zhejiang Province Key Laboratory of Magnetic Materials and Application Technology, Ningbo Institute of Materials Technology and Engineering, Chinese Academy of Sciences, Ningbo, Zhejiang, China.

Soft-magnetic fibers (SMFs) play a crucial role in energy conversion, transmission, and storage within electronic devices. However, conventional SMFs have poor plasticity and are therefore difficult to withstand long-term tensile, torsional, and shear deformation. A high fraction of grain boundaries could improve plastic deformability of conventional SMFs, but deteriorates the coercivity.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!