Halo pin positioning in the temporal bone; parameters for safe halo gravity traction.

Spine Deform

Department of Orthopedic Surgery, University Medical Center Utrecht, Heidelberglaan 100, NL-3584 CX, Utrecht, The Netherlands.

Published: January 2021

Introduction: Halo gravity traction (HGT) is increasingly used pre-operatively in the treatment of children with complex spinal deformities. However, the design of the current halo crowns is not optimal for that purpose. To prevent pin loosening and to avoid visual scars, fixation to the temporal area would be preferable. This study aims to determine whether this area could be safe for positioning HGT pins.

Methods: A custom made traction setup plus three human cadaver skulls were used to determine the most optimal pin location, the resistance to migration and the load to failure on the temporal bone. A custom-made spring-loaded pin with an adjustable axial force was used. For the migration experiment, this pin was positioned at 10 predefined anatomical areas in the temporal region of adult cadaver skulls, with different predefined axial forces. Subsequently traction force was applied and increased until migration occurred. For the load-to-failure experiment, the pin was positioned on the most applicable temporal location on both sides of the skull.

Results: The most optimal position was identified as just antero-cranial to the auricle. The resistance to migration was clearly related to the axial tightening force. With an axial force of only 100 N, which corresponds to a torque of 0.06 Nm (0.5 in-lb), a vertical traction force of at least 200 N was needed for pin migration. A tightening force of 200 N (torque 0.2 Nm or 2 in-lb) was sufficient to resist migration at the maximal applied force of 360 N for all but one of the pins. The load-to-failure experiment showed a failure range of 780-1270 N axial force, which was not obviously related to skull thickness.

Conclusion: The temporal bone area of adult skulls allows axial tightening forces that are well above those needed for HGT in children. The generally applied torque of 0.5 Nm (4 in-lb) which corresponds to about 350 N axial force, appeared well below the failure load of these skulls and much higher than needed for firm fixation.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s43390-020-00194-2DOI Listing

Publication Analysis

Top Keywords

axial force
16
temporal bone
12
force
9
halo gravity
8
gravity traction
8
cadaver skulls
8
resistance migration
8
experiment pin
8
pin positioned
8
traction force
8

Similar Publications

Infrared Ion Spectroscopy of Gaseous [Cu(2,2'-Bipyridine)]: Investigation of Jahn-Teller Elongation Versus Compression.

J Phys Chem A

January 2025

Institute for Molecules and Materials, FELIX Laboratory, Radboud University, Toernooiveld 7, 6525 ED Nijmegen, The Netherlands.

Symmetry breaking is ubiquitous in chemical transformations and affects various physicochemical properties of materials and molecules; Jahn-Teller (JT) distortion of hexa-coordinated transition-metal-ligand complexes falls within this paradigm. An uneven occupancy of degenerate 3d-orbitals forces the complex to adopt an axially elongated or compressed geometry, lowering the symmetry of the system and lifting the degeneracy. Coordination complexes of Cu are known to exhibit axial elongation, while compression is far less common, although this may be due to the lack of rigorous experimental verification.

View Article and Find Full Text PDF

Objective: The aim of this study was to evaluate whether the locking femoral neck plate (LFNP) can be an alternative fixation method to the cannulated screws with a medial buttress plate. For this purpose, we compared biomechanically the LFNP and cannulated screws with or without a medial buttress plate in Pauwels type 3 femoral neck fractures.

Methods: A vertical fracture model was created at an 80-degree angle to the femoral neck in 28 synthetic bone models.

View Article and Find Full Text PDF

Background/purpose: Although clinical studies have suggested a link between non-axial forces and reduced longevity of cervical restorations, the underlying mechanisms require further numerical investigation. This in-silico study employed a cohesive zone model (CZM) to investigate interfacial damage in a cervical restoration subjected to different load directions.

Materials And Methods: A plane strain model of a maxillary premolar was established, with a wedge-shaped buccal cervical restoration.

View Article and Find Full Text PDF

We analyzed the intrinsic strength of distal and proximal FeN bonds and the stiffness of the axial NFeN bond angle in a series of cytochrome b5 proteins isolated from various species, including bacteria, animals, and humans. Ferric and ferrous oxidation states were considered. As assess- ment tool, we employed local vibrational stretching force constants ka(FeN) and bending force constants ka(NFeN) derived from our local mode theory.

View Article and Find Full Text PDF

Background & Objective: Currently, there are many implants in clinical use, making it hard to choose the right one for the patient. The success rate of an implant depends on its diameter, length, and direction of insertion in bone. In implant dentistry, Finite Element Analysis (FEA) simulates intraoral conditions in vitro and analyzes the effects of implant material, diameter, size, and other components related to oral structure on the implant and peri-implant tissues.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!