A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 176

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML

File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 316
Function: require_once

Constancy checks of HDR well-type ionization chambers in the linac era: Potential use of a Sr source. | LitMetric

Constancy checks of HDR well-type ionization chambers in the linac era: Potential use of a Sr source.

Appl Radiat Isot

Radiotherapy Department, University of São Paulo Medical School Hospital and Clinics at Ribeirão Preto, Av. Bandeirantes 3900, 14040-900, Monte Alegre, Ribeirão Preto, SP, Brazil; Department of Physics, Faculty of Philosophy, Sciences and Letters at Ribeirão Preto, University of São Paulo, Av. Bandeirantes 3900, 14040-901, Monte Alegre, Ribeirão Preto, SP, Brazil. Electronic address:

Published: November 2020

AI Article Synopsis

  • A new method using a strontium (Sr) source was adapted for consistency checks of HDR brachytherapy well-type ionization chambers (WTIC).
  • The Sr source was placed on top of the WTIC to establish a reference reading, and measurements were taken to assess variations from this reference.
  • This method was validated against calibrated iridium (Ir) source measurements, suggesting a tolerance limit for variation of 1.5%.

Article Abstract

A methodology using aSr source, originally designed for constancy checks of parallel-plate ionization chambers, was proposed for constancy checks of HDR brachytherapy well-type ionization chamber (WTIC). The Sr source was positioned on the top of the WTIC, a reference reading value was established, and the constancy variation was calculated by the difference between each subsequent measurement and the reference value. This methodology was validated by calibrated Ir sources measurements and an acceptable tolerance limit for constancy variation of 1.5% was suggested.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.apradiso.2020.109309DOI Listing

Publication Analysis

Top Keywords

constancy checks
12
checks hdr
8
well-type ionization
8
ionization chambers
8
constancy variation
8
constancy
5
hdr well-type
4
chambers linac
4
linac era
4
era potential
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!