Objectives: Accurate placement of the pedicle screw is requisite for any successful spinal instrumentation procedure. Screw insertion can be achieved using free-hand and fluoroscopic- or navigation-guided techniques. We sought to assess the variation in accuracy between fluoroscopic- and navigation-guided techniques, which are both used in Sheffield Teaching Hospitals National Health Service Trust, a tertiary spine referral center.
Methods: Using a retrospective study design, we assessed all the pedicle screws placed between 2013 and 2018. Radiographic and clinical assessment of all cases was performed.
Results: We studied 176 spinal instrumented cases, with a total of 831 screws implanted, out of which 296 (35.6%) were navigated and 535 (64.4%) were fluoroscopic guided. Pathology treated included spinal stenosis, spondylolisthesis, tumors, and trauma. Suboptimal screw position was identified in 2.03% (n = 6) of the navigation-guided series and 4.11% (n = 22) of the fluoroscopic-guided series with an overall screw misplacement rate of 3.4%. Evaluating surgeons' individual accuracy rates revealed that suboptimal screw placement registered a higher variation for the fluoroscopy-guided technique, and the misplacement rate was higher for surgeons with a lower volume of cases.
Conclusions: Use of navigation during spinal instrumentation helps lower the rate of screw misplacement for spinal surgeons who are at the beginning of their learning curve or do not frequently perform this kind of procedure. Navigation-guided spinal instrumentation is more accurate compared with fluoroscopic-guided techniques and appears to have a lower complication rate.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.wneu.2020.08.207 | DOI Listing |
Aim: We investigated the short- term results of dynamic/semi-rigid stabilization in patients with cervi-cal spinal stenosis and compare them with patients for which decompression and posterior cer-vical fusion was performed.
Material And Methods: 28 patients were included in this study. Group 1 was the semi-rigid group (four male, ten fe-male), group 2 was the fusion group (nine male, five female).
Front Surg
January 2025
Department of Orthopedic, General Hospital of Ningxia Medical University, Yinchuan, Ningxia, China.
Background: Human brucellosis is the most common bacterial zoonosis worldwide, with brucella spondylitis (BS) being one of its most severe forms, potentially leading to spinal deformity or paralysis. This study aims to provide a comprehensive overview of the current status and research trends in the BS field using bibliometric methods.
Methods: Publications on BS from January 1, 1980, to March 24, 2024, were retrieved from the Web of Science database.
Cureus
December 2024
Orthopaedic and Spine Surgery, National Institute of Traumatology and Orthopaedic Rehabilitation (NITOR), Dhaka, BGD.
Introduction: Ankylosing spondylitis (AS), a chronic inflammatory spondyloarthropathy affecting the spine, progressively leads to increased spinal stiffness. This condition increases the risk of spine fractures in patients, even from trivial injuries. The process of slow bone formation within the ligaments of the spine and the fusion of the spinal diarthrosis contribute to the most prominent symptom of progressive stiffness of joints, predominantly affecting the spine and sacroiliac joints.
View Article and Find Full Text PDFMedicine (Baltimore)
January 2025
Department of Orthopaedics and Traumatology, Faculty of Medicine, Van Yuzuncu Yil University, Van, Turkey.
Anterior cruciate ligament reconstruction aims to improve knee stability and range of motion. The AperFix system consists of polymer components, and fixed-loop fixation is an established endoscopic technique. Our aim in this study was to compare the long-term clinical and radiological results of AperFix and fixed-loop fixation and to prove that the long-term results of the AperFix fixation method are at least as good as those of the fixed loop device.
View Article and Find Full Text PDFEur J Paediatr Neurol
January 2025
Servicio de Neurología Pediátrica, Hospital Universitario La Paz, Madrid, Spain. Electronic address:
Background: The study aimed to describe a new Ommaya reservoir implantation method in late-onset SMA patients, assessing its safety and effectiveness under standard clinical conditions.
Methods: Prospective observational study. Lumbar intrathecal access was unfeasible due to significant scoliosis and prior spinal surgeries with instrumentation.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!