Results of First In Vivo Trial of an Acoustic Self-Tonometer.

Transl Vis Sci Technol

University of Bremen, Bremen Institute for Metrology, Automation and Quality Science (BIMAQ), Bremen, Germany.

Published: August 2020

Purpose: Glaucoma is the world's most common cause of irreversible blindness, which makes early diagnosis, with the goal of preserving vision, essential. The current medical intervention is to reduce intraocular pressure (IOP) to slow down progression of the disease. The main goal of this study was to test a novel handheld acoustic self-tonometer on humans.

Methods: A sound pressure pulse generated by a loudspeaker causes the eye to vibrate. A pressure chamber is placed on the human orbit to form a coupled system comprised of the patient's eye, the enclosed air, and the loudspeaker. A displacement sensor in front of the loudspeaker membrane allows the dynamic behavior of the entire system to be detected.

Results: For this clinical trial series, a prototype of the acoustic self-tonometer principle was applied. The resulting membrane oscillation data showed sensitivity of patient IOP, but direct allocation of the measured damping and frequency to the IOP was not significant. For this reason, an artificial neural network was used to find relationships among the subjects' biometric eye parameters in combination with the self-tonometer data for the IOP reference. An expanded measurement uncertainty ( = 2) equal to 6.53 mm Hg was determined for the self-tonometer in a Bland-Altman analysis using Goldmann applanation tonometer reference measurements.

Conclusions: The usability and success rate of producing valid measurement values with the device during self-measurements by test subjects was nearly 92%. The cross-sensitivities observed require compensation in a possible redesign phase to reduce the measurement uncertainty by at least 25% to the maximum of 5 mm Hg required to seek medical device approval.

Translational Relevance: Building on successful laboratory experiments with pig eyes, this article reports the results of testing the acoustic tonometer on humans.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7442865PMC
http://dx.doi.org/10.1167/tvst.9.9.18DOI Listing

Publication Analysis

Top Keywords

acoustic self-tonometer
12
measurement uncertainty
8
self-tonometer
5
vivo trial
4
acoustic
4
trial acoustic
4
self-tonometer purpose
4
purpose glaucoma
4
glaucoma world's
4
world's common
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!